19
Setting the Laser Wavelength
On the parameter setup page, press [
▲
] or [
▼
] to select Laser wavelength. Press [Enter]
to enter or [ESC] to exit from the laser wavelength setup page, as shown in Figure 3-143.
Figure 3-14 Setting the laser wavelength
Setting the Measurement Mode
Two measurement modes are available: real-time and average.
In real-time mode, the OTDR tests the connection status of the external optical fiber
in real time and update the test trace in time. In real-time mode, the measurement stops
only when you press [Run/Stop]; otherwise, the measurement goes on an on. In average mode,
the OTDR accumulates and averages the measurement data based on the selected measurement
duration. When the measurement time is equal to or longer than the measurement duration
set by the user, the current measurement stops and the measurement result is displayed.
Generally, you can select the average mode for measurement.
On the parameter setup page, press [
▲
] or [
▼
] to select Measurement mode. Press [Enter]
to enter or [ESC] to exit from the measurement mode setup page, as shown in Figure 3-15.
Figure 3-15 Setting the measurement mode
Setting the Length unit
On the parameter setup page, the length unit is set to m by default.
Setting the Reflection Rate