K8853X Air Release Value Instrument
Operation and Instruction Manual
K8853X-Manual
-
5-
4.
Once the top and front panels of the crate
have been removed it will be easy to remove
the 2x4 supports.
5.
Extract the instrument from the crate and
place on suitable cart for transportation to
work area / lab bench.
WARNING:
Be sure two or more individuals
are available for extracting and lifting instrument
from box to cart and from cart to bench.
Individuals must lift in accordance to proper
technique. See Figure below.
6.
Lift instrument from cart and place on lab
bench.
7.
Before opening the Cardboard Box, check the
Shock Watch Label on the Box for indication
of rough handling and possible damage.
8.
Check labeling for correct orientation of
instrument. (e.g. This Side Up)
9.
Carefully open top of box with box cutter and
remove packing foam.
10.
Ensure that all parts listed on the packing list
are present. Inspect the unit and all
accessories for damage. If any damage is
found, keep all packing materials and
immediately report the damage to the carrier.
We will assist you with your claim, if
requested. When submitting a claim for
shipping damage, request that the carrier
inspect the shipping container and equipment.
Do not return goods to Koehler without written
authorization.
3.3 Setup
Equipment Placement.
Make sure the instrument
is placed on a firm, level table. The unit may be
leveled by making minor turning adjustments to
the feet located at the base of the unit. Please
note that Koehler does not supply a level with this
equipment.
Environmental Conditions:
The instrument
environment must comply with the following
conditions for proper setup:
•
No / Low Dust
•
No direct sunlight
•
Not near heating or AC ventilation ducts
•
No Vibrations
•
Clearance from other instruments
•
Temperature Range: 5 to 40°C
•
Elevation to 2000 meters
•
Relative Humidity: < 80%
Glassware.
Place the air inlet tube into the
jacketed test vessel. Secure the assembly to the
front of the main unit, where the rotating plate
serves the resting base for the jacketed test
vessel.
Compressed Air Source.
Connect a source of
oil-free, filtered compressed air source to the main
unit with a pressure of 80-115 psi. The connection
should be made to hose-barb located on the left
side of the main unit.
Power.
Connect the line cords to properly fused
and grounded receptacles with the correct voltage
as indicated in section 1.3 or on the back of the
unit.
WARNING:
For safety, disconnect the
power when performing any maintenance and/or
cleaning.
Do
NOT
turn the power on unless the
bath is filled with the proper medium; otherwise,
damage may occur to the unit and the warranty
will be void.