4.2  Temperature Controller Operation 

 

 

 
 

Figure 4. Temperature Controller 

 

1.  Process  Temperature  Display.

  The  upper 

red  LED  display  shows  the  process 
temperature as read from the RTD probe. 

 

2.  Set  Point Temperature Display.

 The lower 

green  LED  display  shows  the  set  point 
temperature of the controller. 
 

3.  Programming 

Key.

 

Permits 

scrolling 

through  controller  menu  parameters.  One 
Level Forward 
 

4.  Down  Key.

  Used  to  decrease  the  set  point 

temperature  and  to  decrease  or  change 
parameter  values  when  programming  the 
temperature controller.  
 

5.  Up  Key.

  Used  to  increase  the  set  point 

temperature  and  to  increase  or  change 
parameter  values  when  programming  the 
temperature controller. 
 

6.  Exit / Function Key.

 This key is used to exit 

or leave a level. One level backward  
 

IMPORTANT 

NOTE:

 

The 

digital 

temperature  controller  for  the  unit  comes 
pre-programmed  from  the  Koehler  factory. 
Please  do  NOT  attempt  to  re-program  the 
digital temperature controller as this will void 
the  product  warranty.  If  assistance  is 
required,  please  do  not  hesitate  to  contact 
the Koehler technical service department. 

 

Setting  the  Temperature

.  Set  the  desired 

operating  temperature  by  adjusting  the  set 
point with the up and down keys. The set point 
will  be  displayed  in  the  lower  green  Set  Point 
LED display and the actual temperature will be 
displayed  in  the  upper  red  Process  LED 
display.  Please  allow  the  instrument  to  fully 
equilibrate before proceeding with any testing. 

 

Temperature  Calibration.

  This  routine  allows  the 

digital  temperature  controller  to  be  calibrated 
to a certified thermometer.  

 

a.  Use  a  certified  calibrated  measuring 

device 

to 

acquire 

the 

temperature. 

Calculate  the  difference  between  the 
measuring  device  and  the  Process  value 
displayed on the controller. 

b.  Press the program key two times until 

PCt

 

is  displayed  in  the  lower  green  LED 
display.  Press  the  DOWN  key.  CAL  will 
display on the lower green display. If there 
is  a  value  observed  in  the  upper  red  LED 
display,  add  it  to  the  calculated  difference 
obtained  in  the  previous  step.  This  is  the 
offset value. 

 

c.  Press  the  Program  Key.  The  lower  green 

display will flash. Use the up or down keys 
to adjust to the new calibration offset value 
on  the  upper  red  display  calculated  in  the 
previous  step.  When  the  value  has  been 
entered,  the  controller  will  automatically 
store  the  value.  The  lower  green  display 
will stop flashing. If further adjustments are 
necessary,  press  the  Program  Key  again. 
Resume  regular  operations  by  pressing 
the  Exit  /  Function  key  two  times.  Verify  if 
the new calibration is correct by observing 
the  upper  red  display  and  comparing  the 
value with the calibrated reference device.  

 
 

Auto  Tune.

  This  routine  allows  the  digital 

temperature  control  to  learn  the  heating 
parameters needed for any particular set point 
temperature.  This  operation  should  be  done 
when  installing  a  new  unit,  after  replacing  or 
changing  the  bath  medium  type,  or  utilizing  a 
different  temperature  set  point  20%  different 
from 

the 

previously 

used 

set 

point 

temperature. 

 

a.  Set  the  operating  temperature  to  the 

desired setting. 

 

Summary of Contents for K259 Series

Page 1: ... Avenue Bohemia New York 11716 1796 USA Toll Free 1 800 878 9070 US only Tel 1 631 589 3800 Fax 1 631 589 3815 http www koehlerinstrument com e mail info koehlerinstrument com Petroleum Testing Analysis Instrumentation Custom Design Manufacturing K259XX Constant Temperature Water Bath ...

Page 2: ......

Page 3: ...red in conformance with the applicable standards set forth in this certification Specifications ASTM D1657 GPA 2140 This unit is tested before it leaves the factory to ensure total functionality and compliance to the above specifications and ASTM standards Test and inspection records are on file for verification Jesse Kelly Application Engineer Koehler Instrument Company ...

Page 4: ......

Page 5: ...K259XX Constant Temperature Water Bath Operation and Instruction Manual K295XX Manual 1 ...

Page 6: ...K259XX Constant Temperature Water Bath Operation and Instruction Manual K295XX Manual 2 ...

Page 7: ...nformation on subsequent processing or recycling prior to disposal of the equipment http www dpa system dk en WEEE Products Private or professional use Responsibility for Registration and Annual Reporting Koehler will not sell directly to end users in the EU and so has no responsibility to register within each EU state and to make annual reports Koehler declares that this responsibility is born by...

Page 8: ...ely collected WEEE Mercury containing components such as switches or backlighting lamps compact fluorescent lamps CFL Batteries Printed circuit boards if the surface of the printed circuit board is greater than 10 square centimeters about 4 sq in Toner cartridges liquid and pasty as well as color toner Chlorofluorocarbons CFC hydrochlorofluorocarbons HCFC or hydrofluorocarbons HFC hydrocarbons HC ...

Page 9: ...rmation and Warnings 7 3 Getting Started 7 3 1 Packing List 7 3 2 Unpacking 7 3 3 Setup 8 4 Descriptions 9 4 1 Instrument Descriptions 9 4 2 Temperature Controller Operation 11 5 Operation 12 6 Maintenance 12 6 1 Routine Maintenance 12 6 2 Replacement Parts 12 7 Wiring Diagrams 13 7 1 K25900 115V 13 7 2 K25995 230V 50Hz 14 7 3 K25990 230V 60Hz 15 8 Service 16 9 Storage 16 10 Warranty 16 11 Returne...

Page 10: ...for a full range of accurate and reliable instrumentation as well as support for your laboratory testing programs Please do not hesitate to contact us at any time with your inquiries about equipment tests or technical support 1 2 Recommended Resources and Publications 1 American Society for Testing and Materials ASTM 100 Barr Harbor Drive West Conshohocken Pennsylvania 19428 2959 USA Tel 1 610 832...

Page 11: ...hemical Reagents Information The use of chemicals and reagents with this equipment may exhibit potential hazards Any use must be familiarized with the possible dangers before use We recommend consulting the Material Data and Safety Sheet MSDS on each chemical reagent for safety information MSDS information can be easily located on the Internet at http siri uvm edu or http www sigma aldrich com 3 G...

Page 12: ...trument on a firm level surface in an area with adequate environmental control Environmental Conditions The instrument environment must comply with the following conditions for proper setup No Low Dust No direct sunlight Not near heating or AC ventilation ducts No Vibrations Clearance from other instruments Temperature Range 5 to 40 C Elevation to 2000 meters Relative Humidity 80 Filling the Bath ...

Page 13: ...mometer to verify bath temperature displayed on temperature controller 5 Stirrer Motor To agitate bath medium in order to provide temperature uniformity in the apparatus 6 Temperature Measurement Probe Measures direct temperature of bath and communicates with temperature controller for display Figure 2 Instrument Descriptions_Front Panel 7 Line Switch For powered on entire unit Temperature control...

Page 14: ... emptying bath medium to appropriate receptacle 16 Bath Outlet Connect insulated tubing from mechanically refrigerated chiller inlet Figure 6 Item 19 Figure 4 Instrument Descriptions_ Back_115V_230V 60Hz 17 Chiller Power Receptacle_115V For connecting mechanically refrigerated chiller Figure 1 Item 3 power cord to constant temperature bath Figure 1 Item 1 NOTE Plug type for 115V 60Hz and 230V 60Hz...

Page 15: ... to fully equilibrate before proceeding with any testing Temperature Calibration This routine allows the digital temperature controller to be calibrated to a certified thermometer a Use a certified calibrated measuring device to acquire the temperature Calculate the difference between the measuring device and the Process value displayed on the controller b Press the program key two times until PCt...

Page 16: ... Allow time for the bath temperature to equilibrate 7 Proceed with performing the test according to the test methods prescribed in Section 1 2 of this document 6 Maintenance WARNING Disconnect power to the unit before servicing to avoid exposure to high voltages and or temperatures which may result in personal injury or death If you have any questions about maintaining your equipment please do not...

Page 17: ...K259XX Constant Temperature Water Bath Operation and Instruction Manual K295XX Manual 13 7 Wiring Diagrams 7 1 K25900 115V ...

Page 18: ...K259XX Constant Temperature Water Bath Operation and Instruction Manual K295XX Manual 14 7 2 K25995 230V 50Hz ...

Page 19: ...K259XX Constant Temperature Water Bath Operation and Instruction Manual K295XX Manual 15 7 3 K25990 230V 60Hz ...

Page 20: ...he return of the product to the factory The sole responsibility of Koehler Instrument Company and the purchaser s exclusive remedy for any claim arising out of the purchase of any product is the repair or replacement of the product In no event shall the cost of the purchaser s remedy exceed the purchase price nor shall Koehler Instrument Company be liable for any special indirect incidental conseq...

Page 21: ...K259XX Constant Temperature Water Bath Operation and Instruction Manual K295XX Manual 17 Notes ...

Page 22: ...K259XX Constant Temperature Water Bath Operation and Instruction Manual K295XX Manual 18 Notes ...

Reviews: