Kobold ZOE series Operating Manual Download Page 29

  

ZOE 

ZOE K04/0718 

 

Page 29 

Since some useful information cannot be displayed in this case, it is recommended to 
use a bigger unit to avoid the loss or to use the smaller display for totalizers. 

 

9.7  Pulse output (only with ZOE-ZE Supply option 3) 

A scalable pulse output is available for the options ZOE-ZExxxx3xx. When the pulse 
output is enabled, a scaled pulse output train is given at the output to the input pulses. 
The pulse width of the pulses is adjustable between the span of 1 ms to 20 sec. 
When pulse output is activated, the symbol "

" for pulse output is highlighted on the 

display. The electrical output of the pulse output is push-pull type, so HIGH and LOW 
is actively switched through to the output. 
To generate the output pulses, the input pulse train is directly processed in connection 
with the scaling factor “K0”, the output pulses are therefore not derived from the 
calculated volume flow. The pulse output is based on a linear path of the input signal. 
Sensors are used with nonlinear characteristic, with the usage of linearization function 
only the updated volume flow can be linearized, pulse output and volume counter 
cannot be linearized.  
 
Behaviour at OVERFLOW: 
If the measured volumetric flow lies in the OVERFLOW range, the pulse output is 
switched off and a constant HIGH level is switched at the output. 
 
Generation of the output pulse train: 
The maximum adjustable pulse rate for the pulse output is 1000 pulses per liter. This 
means that the minimum pulse volume that can be represented by the pulse output is 
0.001 Liter per pulse. Furthermore, the set pulse volume must meet following 
condition: 
 

 

 

 

∗ 1  

 

100

60 ∗

 

 

0.5

 

 

 

 

In case if the above mentioned condition is not complied with, it may lead to a long 
lagged pulse train at pulse output even when the input frequency signal has been 
interrupted and removed. While configuring the pulse output, if the above condition is 
violated, the user will be informed via notification on the display with “Lagged pulse 
train possible. Please check the settings.” In case if the measuring range endpoint is 
changed during the operation of pulse output, the user needs to check the pulse 
output settings once again. 
The pulse output works only in the measurement mode i.e. in the menu mode, no 
pulses will be generated at the output. The pulses acquired in the menu mode will be 
given out once the normal measuring mode is activated again, leading to no loss but a 
delayed pulse train at the output depending on how long the menu mode remains 
active. 

Summary of Contents for ZOE series

Page 1: ...Operating Manual for Flow Totalizer Model ZOE...

Page 2: ...d Menu Structure 18 8 1 Overall 18 8 2 Measuring Mode 18 8 3 Menu Mode 19 9 Instrument Parameters 21 9 1 End of instrument parameterization 21 9 2 Overview of menu functions instrument parameters 22 9...

Page 3: ...e measuring unit should be used only when the machines fulfil the EC machine guidelines 2 1 Notice according to the Battery Law Old batteries do not belong in household waste You can send the batterie...

Page 4: ...culation display and transfer of calculations and flow rates of flow meters with pulse or frequency outputs The instruments display flow rate day counter resettable and total counter in the operator s...

Page 5: ...n x x Supply voltage DC supply x Battery operation x x Sensor supply only with external supply 8 V Electrical outputs only with external supply Pulse output x LCD Display Selectable units x X Decimal...

Page 6: ...r NPT are available for electric connection in case of electronics with plastic housing To use these cable inlets the existing factory protection caps must be broken with a tool e g screwdriver and a...

Page 7: ...3 6V 2200mAh Size AA ext Batt Service 7 2 Overview of Terminals ZOE ZX with Supply option B and 3 Supply option B only battery either 1 x 3 6V or 2 x 1 5V Supply option 3 with battery either 1 x 3 6V...

Page 8: ...option ZOE ZX can also be powered by a battery In battery mode the device function scope is relatively limited Pulse output is not available The backlight of the display is switched off and cannot be...

Page 9: ...Sensor Inputs 7 4 1 Active pulse signal Power Supply External DC Supply Signal input setting ACTIVE SIGNAL Wiring for ZOE ZE with Supply option 3 Wiring for ZOE Z with Supply option B Battery 1 5V Si...

Page 10: ...ZOE Page 10 ZOE K04 0718 7 4 2 Hall Sensor NPN and PNP Sensors Power supply only external DC supply Signal input setting HALL Signal input setting NPN Signal input setting PNP Wiring for ZOE ZE...

Page 11: ...ZOE ZOE K04 0718 Page 11 7 4 3 Reed switch Signal input setting REED Wiring for ZOE ZE with Supply option 3 Wiring for ZOE Z with Supply option B...

Page 12: ...supply External DC supply and battery operation Battery operation reduces battery life Signal input setting COIL Wiring for ZOE ZE with Supply option 3 The connection of the DOT turbine wheel flowmet...

Page 13: ...ZOE ZOE K04 0718 Page 13 7 4 5 Namur Sensor Power supply only external DC supply Signal input setting NAMUR Wiring for ZOE ZE with Supply option 3 Battery 3 6V 2200mAh Size AA...

Page 14: ...ZOE Page 14 ZOE K04 0718 7 4 6 Transducer model UMF2 Power supply only external DC supply Signal input setting NPN Wiring for ZOE ZE with Supply option 3 5 4 Battery 3 6V 2200mAh Size AA...

Page 15: ...hrough a an active control input or b Carried out by a passive N O contact In both cases the function will be activated on switching from HIGH to LOW level falling edge When using an active control si...

Page 16: ...ZOE Page 16 ZOE K04 0718 Wiring for ZOE ZE with Supply option 3 Battery 3 6V 2200mAh Size AA...

Page 17: ...17 7 6 Pulse Output Pulse output is set to push pull output type and provides an active digital signal with HIGH amplitude of about Vs Wiring for ZOE ZE with Supply option 3 Battery 3 6V 2200mAh Size...

Page 18: ...the supply voltage the electronics starts in measurement mode In this mode the input signals are recorded continuously current instantaneous flow and volume counter readings are periodically calculate...

Page 19: ...mode is turned on the input signal detection and pulse output are still active in the background After leaving the menu mode all display parameters are then updated again in measurement mode Note The...

Page 20: ...all menu groups can not be displayed simultaneously on the screen so the list of menu items can then be scrolled up or down To choose the selected item button should be pressed and the device goes in...

Page 21: ...the cursor can be moved again to left by pressing key and the correction can be made If the cursor is on the far right the set value is saved by pressing the button and the display changes to the pare...

Page 22: ...SIGNAL TYPE nPn For NPN sensors List selection default Battery external COIL For induction coils Battery external PnP For PNP sensors Battery external ACt For active pulse signals Battery external nAm...

Page 23: ...S 10 Battery external COUNTER UNIT mL L m3 galUS galUK User Volume units for counter List selection L Battery external RST PART Reset subset counter A Button selection Yes No No Battery external PULSE...

Page 24: ...to detect even lower frequencies Input Timeout parameter can be configured up to 20 s After the completion of Refresh time the signal processing waits for the full input period until the additional wa...

Page 25: ...automatically activated In case of other sensors it is by default inactive and can be activated by the user anytime using Menu 9 4 Scaling The correct scaling of the signal inputs is necessary for ex...

Page 26: ...ions for the base sequence The sensor curve must be monotonically increasing since the frequency values increase continuously with increasing flow Inconsistent base sequence After entering the lineari...

Page 27: ...ameter start point defines the lower flow rate limit for display and further evaluation If the measured flow rate lies below the measuring range value the flow rate is set to 0 therefore hidden At the...

Page 28: ...totalizer counter values on smaller as well as the bigger display with the help of different layouts available under DISPLAY menu However the bigger display is limited to display only 5 digits If the...

Page 29: ...earized Behaviour at OVERFLOW If the measured volumetric flow lies in the OVERFLOW range the pulse output is switched off and a constant HIGH level is switched at the output Generation of the output p...

Page 30: ...th the activation of a user password the access to menu for the user can be locked by a master user 9 8 1 User service Change password The default administrator user password is set to 00000 the user...

Page 31: ...1 1000 Hz High low threshold 0 6 VDC Minimum input voltage for coil input 30 mVpp Measuring rate 1 s Units for volume flow ml h l m m3 h gpmUK gpmUS gphUK gphUS User Units for volume counter ml l m3...

Page 32: ...sal mount round plastic for retrofitting DON M 3xM20x1 5 cable entry N 3x NPT cable entry F pulse frequency input B 1 battery for operation 0 without 0 without Y special please specify in clear text Z...

Page 33: ...ZOE ZOE K04 0718 Page 33 12 Dimensions mm ZOE ZXK ZOE ZXM...

Page 34: ...2013 Electrical equipment for measurement control and laboratory use EMC requirements Part 1 General requirements Also the following EC guidelines are fulfilled 2014 30 EU EMC Directive 2011 65 EU RoH...

Reviews: