The ZFC does not require special maintenance unless it is used in low-
temperature applications or surroundings with high humidity (above 90% annual
mean). It is the user’s responsibility to take all precautions to dehumidify the
internal atmosphere of the ZFC in such a way that no condensation will occur,
e.g. to put a dose of desiccant (drying agent) inside the enclosure just before
closing it.
Furthermore, it is required to replace the desiccant periodically as advised by its
supplier.
Check periodically:
The condition of the enclosure, cable glands and front panel.
The input/output wiring for reliability and aging symptoms.
The process accuracy. As a result of wear and tear, re-calibration of the
flowmeter might be necessary. Do not forget to re-enter any subsequent K-
Factor alterations.
Clean window and enclosure only with a lint-free cleaning cloth made damp
with a mild soap solution. Do not use any aggressive solvents as these might
damage the coating.
12.2 Instructions for repair
This product cannot be repaired by the user and must be replaced with an
equivalent certified product. Repairs are only allowed to be carried out by the
manufacturer or his authorized agent.