MAS Instruction Manual
Page 19
Figure 4-1
Proper LFE Location
Within The Flow Body
4.2.3
Sensor Tube
CAUTION! Opening the sensor cavity will shift calibra-
CAUTION! Opening the sensor cavity will shift calibra-
CAUTION! Opening the sensor cavity will shift calibra-
CAUTION! Opening the sensor cavity will shift calibra-
CAUTION! Opening the sensor cavity will shift calibra-
tion.
tion.
tion.
tion.
tion.
Do not remove the PCB Bracket (No. 7) unless it is absolutely
necessary to gain access to the sensor cavity. Doing so will shift
the calibration more than 1.5%. The remaining parts of the flow
path are disassembled as shown in the exploded view in Appendix
A. Note the position of the insulation blanket before removal and
re-install in the same fashion. After removal, the sensor tube (No.
5) can be cleaned by purging, washing with a solvent, or by rodding
out the 0.031 inch (.787 mm) internal diameter tube with a 0.029-
0.030 inch (.737-.762 mm) outside diameter rod or wire or with the
Model CK Cleaning Stylet available from Kobold Instruments. To
maximize the time response of your MAS, Kobold has designed
the sensor tube with thin walls. Therefore, when cleaning, be
Therefore, when cleaning, be
Therefore, when cleaning, be
Therefore, when cleaning, be
Therefore, when cleaning, be
extremely careful not to bend the sensor tube or to mar its
extremely careful not to bend the sensor tube or to mar its
extremely careful not to bend the sensor tube or to mar its
extremely careful not to bend the sensor tube or to mar its
extremely careful not to bend the sensor tube or to mar its
inlet or outlet edges.
inlet or outlet edges.
inlet or outlet edges.
inlet or outlet edges.
inlet or outlet edges.
It is important when reinstalling the sensor to make sure that no
torque is imparted on the sensor tube. Torque can be eliminated
by using a good quality oxygen compatible grease on the sensor
sealing “O”-rings. The sensor assembly should slide freely into
the cavity flanges without having to twist it. Twisting will impart
undesirable torque on the sensor and could lead to long term
shifting of the zero value. Also, take E X T R E M E
E X T R E M E
E X T R E M E
E X T R E M E
E X T R E M E care to not
disturb or unravel the sensor windings.
.61
(15.49)
.60
(15.24)
INLET
OUTLET
.85
(21.59)
.90
(22.86)
DIMENSIONS IN INCHES
(DIMENSIONS IN MILLIMETERS IN PARENTHESES)
Summary of Contents for MAS Series
Page 27: ...MAS Instruction Manual Page 26 APPENDIX A Exploded View of MAS Transducers and Parts Lists...
Page 28: ...MAS Instruction Manual Page 27 Aluminum Assembly High Flow MAS 1100 2100 M82 0022...
Page 30: ...MAS Instruction Manual Page 29 MAS Assembly MAS 1000 2000 M82 0023...
Page 32: ...MAS Instruction Manual Page 31 Electronics Enclosure Sub Assembly MAS M82 0019 XXXX B...
Page 34: ...MAS Instruction Manual Page 33 Sensor Compartment Assembly MAS M82 0020 X B...
Page 47: ...MAS Instruction Manual Page 46...
Page 54: ...MAS Instruction Manual Page 54...