58
13 Commissioning
Commissioning
1.
Check air and water connection
!
CAUTION
Make sure that air and water have not been interchanged.
2.
Check media connection.
3.
Check electrical connections.
4.
Switch on compressed air.
5.
Adjust pressure according to the air pressure gauge using the pressure
regulator.
6.
Check tightness: When the compressed air is shut off directly at the Unical,
pressure may decrease by max. 10 % within 30 sec.
7.
Turn water supply on.
8.
Check pressure at water pressure gauge, check tightness.
9.
Switch on power supply for Protos.
!
CAUTION
Only operate the retractable fitting when a sensor is installed!
When the sensor has been removed, it must always be replaced by a dummy!
SensoGate WA 130(X)
Turn the SensoLock ring to “lock“ position to prevent that the retractable
fitting is accidentally moved into the process. This is only possible with the
retractable fitting in “Service“ position (see SensoGate manual).