Removing a Gel-Electrolyte Sensor
Short immersion depth
Removing the Sensor
1) The sensor must only be removed in
SERVICE position
.
2) Remove protective cap (ZU 0759) if provided.
3) Remove cable socket with cable
(G)
.
4) Before removing the sensor, check that there is
no liquid leaking from the outlet (process sealing
might be defective).
5) Remove the sensor (recommended tool: 19 mm,
e.g. Knick ZU0647 wrench).
6) Check whether slide washer
(C)
or O-ring
(D)
(or HD gasket) on the sensor are missing or
damaged.
CAUTION!
When replacing damaged sensors (glass break-
age), you must check the sensor seal in the
immersion tube and replace it if required.
(See section “Replacing the Immersion Tube” on
A
B
C
D
E
F
G