Installing and Removing a Sensor
SensoGate® WA 131 H
Sensors must only be installed or removed by trained personnel authorized by the
operating company.
Make sure that the sensor lock-gate is in SERVICE position
(see “Function Description” on page 10)
.
WARNING!
Process fluids leaking from the outlet or at the leakage holes (see “Build-Up of the
Sensor Lock-Gate” on page 13) show that the calibration chamber is not tight.
Be sure to follow the assembly instructions step by step.
Preparations:
• Check whether the sensor is damaged (glass broken?).
Never install a damaged sensor.
• Check whether slide washer or O-ring on the sensor are damaged and replace if required.
• Remove watering cap from the sensor tip and rinse sensor with water.
• Internally pressurized sensors might have a silicone seal on the diaphragm (as transport
protection). Remove this seal using the knife shipped with the sensor.