Maintenance Work on the Drive Unit
SensoGate® WA 131 H
The drive unit must be removed, for example:
• for general maintenance or inspection
• to clean the calibration chamber, e.g. after a sensor has broken
• to change the sensor / calibration-chamber gaskets
• in the event of a technical fault of the drive unit.
WARNING!
To separate the sensor lock-gate safely from the process, make sure
that it is disconnected from all process media and process pressure.
CAUTION!
Before working on the drive unit, make sure that the sensor lock-gate is in
SERVICE position (see “Function Description” on page 10).
Indication of SERVICE or PROCESS Position
Service position – indicated by the
sensor connector
(A)
protruding
out of the drive unit.
Service position – indicated by
the rubber bellows
(A)
being
expanded
Short immersion depth
Solid-electrolyte sensor
Short immersion depth
Liquid-electrolyte sensor
A
A
SERVICE
PROCESS
SERVICE
PROCESS