55
pH
Oxy Cond
ORP
Apply
Cancel
Calibration
Enter the reference
temperature.
24.4 °C
Temp. offset
Reference temperature
-0.3 K
24.1 °C
Enabling the Option 002 Temp.Cal
(included in Option 001 SOP)
1) In measuring mode, press the Menu softkey.
2) Select “Configuration” and confirm by pressing enter.
3) Select Option “002 Temp.cal” and enter your activation code.
Selecting Temperature Calibration
1) In measuring mode, press the Menu softkey.
2) Select “Calibration” and confirm by pressing enter.
3) Select the “Temperature” calibration mode and confirm by pressing enter.
For Memosens sensors, you can perform a 1-point calibration of the internal
temperature detector. To do so, enter the reference temperature and confirm the
temperature adjustment by pressing the Apply softkey:
Options