background image

7

Documentation

CD-ROM

Complete documentation:
•  User manuals in different languages
•  Safety instructions 
•  Quickstart guides

Safety Instructions

In official EU languages and others.
•  EU Declarations of Conformity

Quickstart Guides

Installation and first steps:
•  Operation
•  Menu structure
•  Calibration
•  Error messages and recommended actions
 
Various languages on CD-ROM and on our website:  
www.knick.de

Specific Test Report

Summary of Contents for Portavo 908 MULTI

Page 1: ...10 ATEX E089 X II 1G Ex ia IIC T3 T4 T6 SE 706X 2 NMSN 14163 Berlin BVS 10 ATEX E089 X II 1G Ex ia IIC T3 T4 T6 II 1G Ex ia IIC T3 T4 T6 SE 706X 1 NM 14163 Berlin BVS 10 ATEX E II 1G Ex ia IIC SE 706X 2 NM 14163 Berlin BVS 10 ATEX E II 1G Ex ia IIC x ia IIC T3 T4 T6 ATEX E 089 X SE 706X 2 NMS 14163 Berlin BVS 10 ATEX E0 II 1G Ex ia IIC T Latest Product Information www knick de Portavo 908 MULTI Us...

Page 2: ......

Page 3: ...fore shipment In that case please attach a corresponding certificate for the health and safety of our service personnel Disposal Please observe the applicable local or national regulations concerning the disposal of waste electrical and electronic equipment Registered Trademarks The following names are registered trademarks For practical reasons they are shown without trademark symbol in this manu...

Page 4: ... and Oxy only 16 Sensor Monitor 17 Messages 17 MemoLog Memosens only 17 Device Info 18 Device Test 18 pH Configuration 19 ORP Configuration 21 Conductivity Configuration 23 Oxygen Configuration 26 pH Calibration 28 Calimatic Calibration 28 Manual Calibration 29 Data Entry Calibration 29 Calibrating a pH ORP Combo Sensor 29 ORP Calibration 30 ISFET Calibration 31 Conductivity Calibration 32 Auto Ca...

Page 5: ...Life 42 Starting Stopping the Data Logger 44 Viewing the Logger Data 44 Deleting the Logger Data 44 Paraly SW 112 Software 46 Error and Status Messages 47 Sensoface Messages 48 Info and Help Texts 48 Error Messages 49 Option 001 SOP 50 Option 002 Temp Cal 50 How to Enter the Rescue PIN Code 54 pH Product Line 56 pH Sensors 56 Knick CaliMat pH Buffer Solutions 57 Accessories for pH ORP 57 Conductiv...

Page 6: ...s Portavo 908 MULTI Meter incl 4 batteries AA and premounted quiver x Carrying strap x Data carrier with detailed user manuals and Paraly SW 112 software x USB cable 1 5 m x Safety instructions x Quickstart instructions in various languages x Specific test report x USB adapter A female to B male for printer connection x ...

Page 7: ...ickstart guides Safety Instructions In official EU languages and others EU Declarations of Conformity Quickstart Guides Installation and first steps Operation Menu structure Calibration Error messages and recommended actions Various languages on CD ROM and on our website www knick de Specific Test Report ...

Page 8: ...iver protects the sensor and prevents it from drying out Furthermore it can be used for calibration The rugged housing is made of a high performance polymer It provides high impact resistance and dimensional stability even when exposed to extreme moisture Scratch proof clear glass display perfectly readable even after years Long operating time with one set of batteries 4 x AA or use of a Li ion ba...

Page 9: ...emosens capable device Sensoface Sensoface provides quick information on the sensor condition The three smiley faces as shown on the right represent the sensor condition during measurement and after a calibration When the condition deteriorates a status message gives a hint to the cause Calimatic pH Calimatic is a very convenient method for pH calibration with automatic buffer recognition You only...

Page 10: ...back for operation Hook A fold out hook on the back allows the meter to be suspended This leaves your hands free for the actual measurement The rating plate is located beneath the hook Protective Cover and Hook Combined Cover and hook can be joined together to form a benchtop stand allowing com fortable and fatigue free working at a lab bench or desk ...

Page 11: ... in the display above the key Arrow keys Selecting Adjusting entries enter Confirming an adjustment on off Switching on off meas Switching on Immediate access of measuring mode Toggling the display Displaying time and date Menu selection Information Return Menu Icons Information Calibration Data logger Configuration Start logger ...

Page 12: ...fails to perform the intended function prolonged storage at temperatures above 70 C 158 F severe transport stresses In this case a professional routine test must be performed This test should be carried out at our factory A special lithium ion battery ZU 0925 suited to the battery compartment is avail able for the Portavo 908 Only this battery type can be charged directly from the USB port A batte...

Page 13: ...parate Temperature Probe After power on a separate temperature probe is automatically recognized If you want to replace the temperature probe you must switch off the meter and then switch it on again Start Up Memosens sensors have a cable coupling which allows convenient replacement of sensors while the cable remains connected to the meter The connecting cable is connected to socket b or e Connect...

Page 14: ...itches to measuring mode With on off first a self test is performed and then the calibration data and settings are displayed before the meter switches to measuring mode Error message Battery charging level Sensor condition Memosens sensor Temperature Icons Important information about the state of the device Left softkey Measured variable Right softkey Start Up ...

Page 15: ...g on the respective sensor pH pH ORP Cond Oxy ISFET ORP Optical Oxy Manufacturer x x x x x x Ref No x x x x x x Sensor serial no x x x x x x Membrane serial no x TAG x x x x x SW version x x x x x x HW version x x x x x Calibration x x x x x x Zero point x x x Slope x x x x ORP calibration x Correction x Nom cell constant x Temp offset x x x x Sensor operating time x x x x x x Membrane operating t...

Page 16: ...time Calibration timer Wear Memosens Parameters which cannot be checked are shown as inactive gray and are set to 100 The parameter values should lie between the outer 100 and inner 50 hexa gon When a value enters the inner hexagon 50 the corresponding caption text flashes red see example Information Help Return Network diagram pH Response time Sensocheck Zero point Cal timer Wear Slope Example Ne...

Page 17: ... in the configuration menu With MemoLog activated up to 100 calibration records can be directly saved to the device After every calibration the complete Memosens data will be recorded Convenient management of the calibration data is possible using the MemoSuite or Paraly SW 112 software MemoLog is not suitable for SE 340 optical oxygen sensor Information Sensor Monitor Shows the raw values availab...

Page 18: ...isted below A green checkmark shows that the test was successful FLASH program memory FLASH data memory FLASH parameter memory RAM working memory Display Test 1 Select Display test and press enter 2 The display lights up red green blue and then white 3 Press any key to stop the test Keypad Test 1 Select Keypad test and press enter 2 Press all nine keys one after the other A green checkmark shows t...

Page 19: ... 00 pH 0 000 pH Wear On Off Calibration Cal timer Off On Interval On 00 99 days Cal mode Calimatic Manual Data entry Cal points Auto 1 point 2 point 3 point Buffer set Mettler Toledo Knick CaliMat Ciba NIST technical NIST standard Hach WTW Hamilton Reagecon DIN 19267 User buffer 1 2 00 4 01 7 00 9 21 2 00 4 00 7 00 9 00 12 00 2 06 4 00 7 00 10 00 1 68 4 00 7 00 10 01 12 46 1 679 4 006 6 865 9 180 ...

Page 20: ... 1 0 Delta mV 0 2000 mV 1 mV 2nd diff On Off Delta C 0 99 9 C 1 0 C Delta F 0 450 F 1 0 F Intv Diff Interval See logger type Interval Difference See logger type Difference Limit value Interval Basis Event 00 00 01 12 59 59 00 01 00 00 00 01 Limit values Min Max corresponding to permissible range see Specifications Printer Driver Auto HP PCL Epson Samsung IBM Paper A4 Small Options 001 SOP 002 Temp...

Page 21: ...ol Italiano Français Português Русский Auto off Off 5 min 10 min 30 min 60 min Temperature C F Right softkey Logger Start Stop Hold value Print Calibration MemoLog Off On TAG Off On Time Date Time format 24 h 12 h Date format dd mm yyyy yyyy mm dd dd mm yyyy mm dd yyyy Time hh mm ss Date Date format as configured Display Appearance Modern Retro Lighting Permanent 60 min 30 min 10 min 5 min 1 min 3...

Page 22: ...terval Difference See logger type Difference Limit value Interval Basis Event 00 00 01 12 59 59 00 01 00 00 00 01 Limit values Min Max corresponding to permissible range see Specifications Printer Driver Auto HP PCL Epson Samsung IBM Paper A4 Small Options 001 SOP 002 Temp cal Add on function enabled via TAN Factory setting Yes No Note Reset to factory settings will also erase all logger data Conf...

Page 23: ...10 min 30 min 60 min Temperature C F Right softkey Logger Start Stop Hold value Print Cond sensor Conductivity S cm S m Range selection Auto o ooo µS cm oo oo µS cm ooo o µS cm oooo µS cm oo oo mS cm ooo o mS cm oooo mS cm Compensation Off MΩ cm TC SAL TDS by wt TC comp TC Linear NLF NaCl HCl NH3 NaOH TC of solution TC 0 20 0 K 2 1 K Ref temp TC 0 100 0 C 25 C 32 212 F 77 F TDS factor TDS 0 9 99 1...

Page 24: ...lation factor Zero point Cal solution NaCl 0 01 mol l NaCl 0 1 mol l NaCl sat KCl 0 01 mol l KCl 0 1 mol l KCl 1 mol l MemoLog Off On TAG Off On Time Date Time format 24 h 12 h Date format dd mm yyyy yyyy mm dd dd mm yyyy mm dd yyyy Time hh mm ss Date Date format as configured Display Appearance Modern Retro Lighting Permanent 60 min 30 min 10 min 5 min 1 min 30 sec Brightness Bright Standard Dim ...

Page 25: ...lta F 0 450 F 1 0 F Intv Diff Interval See logger type Interval Difference See logger type Difference Limit value Interval Basis Event 00 00 01 12 59 59 00 01 00 00 00 01 Limit values Min Max corresponding to permissible range see Specifications Printer Driver Auto HP PCL Epson Samsung IBM Paper A4 Small Options 001 SOP 002 Temp cal Add on function enabled via TAN Factory setting Yes No Note Reset...

Page 26: ...ero point Data entry Calibration timer Off On Interval On 0 99 days MemoLog Off On TAG Off On Time Date Time format 24 h 12 h Date format dd mm yyyy yyyy mm dd dd mm yyyy mm dd yyyy Time hh mm ss Date Date format as configured Oxygen Configuration 1 In measuring mode press the Menu softkey 2 Select Configuration and confirm by pressing enter 3 Make the desired adjustments The following table gives...

Page 27: ...9 99 mbar 1 mbar 2nd diff On Off Delta C 0 99 9 C 1 0 C Delta F 0 450 F 1 0 F Intv Diff Interval See logger type Interval Difference See logger type Difference Limit value Interval Basis Event 00 00 01 12 59 59 00 01 00 00 00 01 Limit values Min Max corresponding to permissible range see Specifications Printer Driver Auto HP PCL Epson Samsung IBM Paper A4 Small Options 001 SOP 002 Temp cal Add on ...

Page 28: ...onfigurable via Paraly SW 112 software 2 Immerse the sensor in the 1st 2nd 3rd buffer solution and press Continue repeat this step for each calibration point 3 Finally the calibration data will be displayed You can Apply or Discard these values Note To abort calibration you can press meas at any time pH Calibration 1 In measuring mode press the Menu softkey 2 Select Calibration and confirm by pres...

Page 29: ...ting calibration see page 36 3 Finally the calibration data will be displayed You can Apply or Discard these values Data Entry Calibration Calibration by entering known sensor values 1 Press the Start softkey 2 Enter the known sensor values for zero and slope 3 Finally you can Apply these values or Cancel the calibration Calibration Note To abort calibration you can press meas at any time Calibrat...

Page 30: ...o do so set TAG to On in the configuration menu default setting Off 5 Enter the temperature corrected setpoint of the calibration solution 6 Immerse the sensor in the calibration solution and wait until the measured value is stable 7 Apply or Discard the ORP setpoint Note Calibration is not possible when the device is connected via USB with the Paraly SW 112 software Calibration Note To abort cali...

Page 31: ...for the first sensor calibration Calibration mode Calimatic Manual Data entry ISFET zero operating point 2 Press the Start softkey 3 Adjust the buffer value if required default pH 7 00 4 Press the Start softkey 5 Finally you can Apply or Discard the calibration value for the operating point When you apply the calibration value the operating point will be stored in the device but not in the sensor ...

Page 32: ...1 mol l KCl 0 1 mol l KCl 1 mol l 2 Press the Start softkey 3 Immerse the sensor in the solution and press Continue 4 Finally the calibration data record will be displayed You can Apply or Discard these values Note Calibration is not possible when the device is connected via USB with the Paraly SW 112 software Conductivity Calibration 1 In measuring mode press the Menu softkey 2 Select Calibration...

Page 33: ...d conductivity value is reached Then press enter 4 Finally you can Apply these values or Cancel the calibration Contacting conductivity sensor conductive Cell constant SE 215 MS 1 00 cm 2 Toroidal conductivity sensor inductive Cell factor SE 680 MS 6 4 cm Zero Calibration 1 Make sure that the sensor is outside the medium in air 2 Press the Start softkey 3 Finally you can Apply these values or Canc...

Page 34: ...G to be edited To do so set TAG to On in the configuration menu default setting Off 6 Perform the selected calibration as described on the following pages Follow the instructions on the display Calibration in Air Calibrating the slope in air 1 Place sensor in air and wait for a stable measured value 2 Press the Start softkey 3 Adjust the correct value for Relative humidity Then press Continue Cali...

Page 35: ...oftkey Calibration will be performed 3 Finally you can Apply these values or Cancel the calibration Data Entry Calibration Calibration by entering known sensor values 1 Press the Start softkey 2 Adjust the known sensor values for zero and slope 3 Finally you can Apply these values or Cancel the calibration Calibration Note To abort calibration you can press meas at any time ...

Page 36: ...ured 4 Watch the display and wait for the reading to stabilize Note Measurement can also be controlled via the Paraly SW 112 software Toggling the Measured Value Display During measurement you can toggle between display of primary secondary measured values and clock by pressing meas Adjusting the Temperature When you connect a sensor without temperature detector you can manually adjust the tempera...

Page 37: ...ter and the paper size Printer Driver Auto HP PCL Epson Samsung IBM Paper A4 Small Printing Currently Measured Values 1 Use the included USB adapter A female to B male to connect a printer to the micro USB port of the meter and switch it on 2 Switch the meter on using the on off or meas key 3 Press the right softkey Print The currently measured values will be printed Note A4 printers do not eject ...

Page 38: ...u will receive a printout of the calibration record with date and time stamp example Printing calibration record Date 04 01 2016 09 19 Device 1nLormac1on Manufacturer Serial no Device type SW version Sensor information Sensor model Manufacturer Order no Serial no SW version li W version TAG Temp offset Ot erating time Wear SIP Calibration data Date of calibration Zero point Slope Buffer 1 Buffer 2...

Page 39: ...and threshold controlled logging The data logger records up to 10 000 entries which can be assigned to different points of measurement TAGs and notes The following data will be recorded meas point note sensor ID serial number of sensor Memosens primary value temperature time stamp device status It is always the currently selected process variable which is recorded Data logger activated Number of m...

Page 40: ...and then save it Difference When the delta range process variable and or temperature related to the last entry is exceeded a new entry is created and the delta range is displaced upwards or downwards by the delta value The first entry is automatically created when the data logger is started Example First entry 7 Delta 1 Interval Time Controlled In the Interval mode the data are cyclically recorded...

Page 41: ...d value remains within the delta range logging is performed at the preset interval The first DIFF entry is automatically created when the data logger is started Limit Value Combined When one of the two limit values Min Max is exceeded the data are logged as defined by the event interval Additionally the last ten measured values before an event are recorded pre trigger As long as the measured value...

Page 42: ...ta Logger Configuring the Data Logger Prerequisite Data logger is stopped The Data logger menu shows the number of occupied entries as well as the number of free entries Configuration can also be done in the Configuration menu under Data logger 1 Press Menu softkey 2 Select Data logger and confirm by pressing enter 3 Select Configure data logger and confirm by pressing enter 4 Configure data logge...

Page 43: ... wt 1 Delta MΩcm 0 9 999 MΩcm 1 0 MΩcm Delta salinity 0 00 45 0 g kg 1 0 g kg Delta TDS 0 00 2000 0 mg l 1 mg l Delta saturation 0 200 Air 1 Air Delta conc 0 20 0 mg l 1 mg l Delta mbar 0 1000 mbar 1 mbar 2nd difference On Off Delta C 0 99 9 C 1 0 C Delta F 0 450 F 1 0 F Intv Diff Interval See logger type Interval Difference See logger type Difference Limit value Interval Basis 00 00 01 12 59 59 1...

Page 44: ...entries proceed as follows 1 Press Menu softkey 2 Select Data logger using the arrows and confirm by pressing enter 3 Select Delete logger data using the arrows and confirm by pressing enter 4 Select deletion mode Complete Data Meas point or Filter you can filter for measuring point parameter or time 5 Press Delete softkey The data are deleted according to the configuration 6 Press Back softkey to...

Page 45: ...data Measured temperature Date and time Primary mea sured value Scroll using arrow keys Example 2 Curve characteristic Entry 2 of 3 Select entry using arrows Data Logger Meas point ABC Note Sensor Analog Data Logger Return Graphic Time Return Measured values Date 45 pH Oxy Cond ORP ...

Page 46: ...avo USB port is connected to the computer The Paraly SW 112 software stands out by the following features Intuitive Windows user interface Easy configuration and management of several meters Display of device and sensor information Configuration of individual buffer sets pH Convenient management and evaluation of the data logger Export function for Microsoft Excel Print function Updating the devic...

Page 47: ...ey icon provides information on the sensor condition main tenance request Measurement can still be performed After a calibration the corre sponding Sensoface icon friendly neutral sad is shown together with the calibration data Otherwise Sensoface is only visible in measuring mode Help text for error 21 Example of an error message Press enter and Help to access the help text Hold value Menu No sen...

Page 48: ...ss impedance Info 05 Zero Slope Info 06 Response time too long Info 07 Operating point ISFET Info 08 Leakage current ISFET Info 09 ORP offset Info 10 Polarization Info and Help Texts When an error or status message appears on the screen proceed as follows to view the corresponding info or help text 1 Press enter 2 Press the Help softkey 3 The help text will be displayed In most cases you can remed...

Page 49: ...al buffers Use a buffer solution with a different nominal value before starting the next calibration step ERR 10 Buffers interchanged Repeat calibration ERR 11 Unstable value Drift too high Leave the sensor in the liquid until the measured value is stable If this does not help replace the sensor ERR 14 Time and date invalid Set the date and time ERR 18 System error Restart reset to factory setting...

Page 50: ...ferent access rights for configuration or calibration Sensor Verification To make sure that only selected sensors can be operated on the meter you can evaluate the sensor type and or the TAG and Group data stored in the sensor The sensor will only be accepted if the data stored in the sensor corresponds to the data stored in the meter Temperature Adjustment For Memosens sensors you can perform a 1...

Page 51: ...1 point 2 point 3 point Buffer 1 Buffer set Mettler Toledo Knick CaliMat Ciba NIST technical NIST standard Hach WTW Hamilton Reagecon DIN 19267 User buffer 1 2 00 4 01 7 00 9 21 2 00 4 00 7 00 9 00 12 00 2 06 4 00 7 00 10 00 1 68 4 00 7 00 10 01 12 46 1 679 4 006 6 865 9 180 4 01 7 00 10 01 12 00 2 00 4 01 7 00 10 00 2 00 4 01 7 00 10 01 12 00 2 00 4 00 7 00 9 00 12 00 1 09 4 65 6 79 9 23 12 75 Bu...

Page 52: ...ion After each calibration step the identified buffer is marked off in the display The next operation procedure is displayed Perform the calibration following the instructions given in the display Continue Return Calibration pH 4 06 Dip sensor in 1st buffer 24 1 C First calibration step completed pH 4 06 Continue Calibration pH 7 61 Dip sensor in 2nd buffer 24 1 C pH 7 00 pH 9 21 pH 4 06 pH 7 00 p...

Page 53: ...er User 3 User 4 Continue Return Configuration Management Disable User 1 ADMIN PIN code cal level conf level 1989 Access Access Creating a User Changing the PIN Code 1 Select a user e g User 1 default ADMIN PIN code 1989 Continue Return Configuration Management Disable User 1 ADMIN PIN code cal level conf level 1989 Access Access 2 Select ADMIN to open the editor for entering the user name enter F...

Page 54: ...When a wrong sensor is connected an error message will be displayed Nevertheless you can continue working with the sensor Reject Here you specify values with which the sensor will be rejected Sensor verification Check model Off Info Reject Check TAG Off Info Reject Check group Off Info Reject How to Enter the Rescue PIN Code 1 In measuring mode press the Menu softkey 2 Select Configuration using t...

Page 55: ... enter 3 Select Option 002 Temp cal and enter your activation code Selecting Temperature Calibration 1 In measuring mode press the Menu softkey 2 Select Calibration and confirm by pressing enter 3 Select the Temperature calibration mode and confirm by pressing enter For Memosens sensors you can perform a 1 point calibration of the internal temperature detector To do so enter the reference temperat...

Page 56: ...nnect other Memosens sensors for pH or ORP as well as the SE 5xx Series Memosens pH ORP combo sensors Temperature Detectors Ref No Pt1000 temperature detector ZU 6959 Pt1000 temperature detector with tilted tip ZU 0156 Memosens sensors have a cable coupling which allows convenient replacement of sensors while the cable remains connected to the meter pH Product Line ...

Page 57: ...0 12 00 0 05 250 ml CS P1200 250 Buffer Sets Set 4 00 3 x 250 ml CS PSET4 Set 7 00 3 x 250 ml CS PSET7 Set 9 00 3 x 250 ml CS PSET9 Set 4 00 7 00 9 00 250 ml each CS PSET479 Accessories for pH ORP Item Order No Robust field case for meter sensor various small parts and user manual ZU 0934 Replacement quiver 5 units ZU 0929 Flexible Memosens lab cable M8 4 pin CA MS 001XFA L Flexible connecting cab...

Page 58: ... SE 615 1 MS You can also connect other Memosens sensors of the SE 6xx Series Temperature Detectors Pt1000 temperature detector ZU 6959 Pt1000 temperature detector with tilted tip ZU 0156 Memosens sensors have a cable coupling which allows convenient replacement of sensors while the cable remains connected to the meter Conductivity Product Line ...

Page 59: ...ting cable for SE 680 sensor M12 4 pin M8 4 pin CA M12 001M8 L Li ion battery ZU 0925 Please visit our website for more information on our product range www knick de Conductivity Standards for determining a cell constant Ready to Use Solutions Quantity Ref No 15 µS cm 0 0001 mol l KCl 300 ml ZU 0350 147 µS cm 0 001 mol l KCl 500 ml ZU 0702 1413 µS cm 0 01 mol l KCl 250 ml ZU 0349 12 88 mS cm 0 1 m...

Page 60: ...ilted tip ZU 0156 Accessories for Oxygen Item Ref No Robust field case for meter sensor various small parts and user manual ZU 0934 Replacement quiver 5 units ZU 0929 Flexible Memosens lab cable M8 4 pin CA MS 001XFA L Flexible connecting cable for Memosens sensors M12 8 pin CA MS 001XDA L Li ion battery ZU 0925 O2 membrane kit 4 x membrane module O ring set 25 ml electrolyte ZU 0564 O2 electrolyt...

Page 61: ...isplay QVGA TFT display with white backlighting Keypad on off meas enter 2 context sensitive softkeys Data logger 10 000 memory locations Recording Manual interval or event controlled with management of tag numbers and notes Cal data logger MemoLog Memosens only Up to 100 Memosens calibration records can be saved Recording Directly retrievable via MemoSuite or Paraly SW 112 USB Viewable on the dis...

Page 62: ...ation data 10 years EMC EN 61326 1 General Requirements Emitted interference Class B residential environment Immunity to interference Industrial environment EN 61326 2 3 Particular Requirements for Transmitters RoHS conformity According to directive 2011 65 EU Power supply 4 x AA alkaline batteries or 1x Li ion battery USB chargeable Nominal operating conditions Ambient temperature 10 55 C 14 131 ...

Page 63: ...00 4 00 7 00 9 00 12 00 03 Ciba 94 2 06 4 00 7 00 10 00 04 NIST technical 1 68 4 00 7 00 10 01 12 46 05 NIST standard 1 679 4 006 6 865 9 180 06 HACH 4 01 7 00 10 01 12 00 07 WTW techn buffers 2 00 4 01 7 00 10 00 08 Hamilton 2 00 4 01 7 00 10 01 12 00 09 Reagecon 2 00 4 00 7 00 9 00 12 00 10 DIN 19267 1 09 4 65 6 79 9 23 12 75 U1 User loadable via Paraly SW 112 Permissible calibration range Zero ...

Page 64: ...M8 socket 4 pins for flexible Memosens lab cable or M12 socket 8 pins for flexible connecting cable for Memosens sensors Display ranges 4 mV 1999 1999 mV Temperature 50 250 C 58 482 F Sensor standardization ORP calibration zero adjustment Permissible calibration range ΔmV offset 700 700 mV ...

Page 65: ... 0 1999 mg l 10 40 C 50 104 F Concentration 0 00 100 by wt Concentration determination NaCl 0 26 wt 0 C 32 F 0 28 wt 100 C 212 F HCl 0 18 wt 20 C 4 F 0 18 wt 50 C 122 F NaOH 0 13 wt 0 C 32 F 0 24 wt 100 C 212 F H2 SO4 0 26 wt 17 C 1 4 F 0 37 wt 110 C 230 F HNO3 0 30 wt 20 C 4 F 0 30 wt 50 C 122 F H2 SO4 94 99 wt 17 C 1 4 F 89 99 wt 115 C 239 F HCl 22 39 wt 20 C 4 F 22 39 wt 50 C 122 F HNO3 35 96 w...

Page 66: ...ning sponge Input optical oxygen M12 socket 8 pins OXY meas ranges at 20 C 68 F Saturation 0 000 200 0 Concentration 000 μg l 20 00 mg l Partial pressure 0 0 1000 mbar Response time t90 30 s t99 60 s Measurement error1 2 3 Zero signal 0 1 full saturation Temperature meas range 4 0 50 C 32 122 F Measurement error1 2 3 Temperature 0 2 K Sensor standardization Automatic calibration in air Zero calibr...

Page 67: ...ductivity calibration 32 Automatic calibration conductivity 32 Automatic calibration pH 28 Automatic self test 18 Auto pH calibration 28 B Basic interval limit value logger type 41 Battery capacity 12 Battery compartment 12 Battery icon 12 Battery life increasing 42 Battery replacement 12 Benchtop stand 10 Buffer set pH configuration 19 Buffer sets 57 Buffer sets configuring pH 46 Buffer solutions...

Page 68: ... points pH 28 Calibration record 15 Calibration record printout 38 Calibration SOP Cal Option 50 Calibration Temp cal Option 55 Calibration zero point operating point ISFET 31 CaliMat buffer solutions 57 Calimatic calibration 28 Calimatic pH 9 Cal mode 31 CD ROM 7 Cell constant conductivity calibration 33 Charge level of batteries 12 CIP sensor information 15 Clearing the logger 44 Combo sensor pH...

Page 69: ...5 Data input pH calibration 29 Data logger configuration 42 Data logger icons 39 Data logger operating modes 40 Data logger starting 44 Data logger stopping 44 Data of the meter 61 Deleting the logger data 44 Delta range data logger 40 Device info Information menu 18 Device properties 8 Device test Information menu 18 Difference logger type 40 Digital conductivity sensors 58 Digital optical oxygen...

Page 70: ...ures 8 Field case accessory 57 59 60 H Help texts 48 Hook 10 I Icons for data logger 39 Icons menus 11 Icons status messages 14 In air Oxy calibration 34 Increasing the battery life 42 Individual buffer sets pH 46 Information menu 15 Info texts 48 Inserting the batteries 12 Installation factor calibration 33 Interval and difference logger type 41 Interval logger type 40 Introduction 8 ISFET calibr...

Page 71: ...IN code 53 M Manual calibration pH 29 meas switch on 14 Measured values printout 37 Measuring 36 Membrane kit accessory 60 Membrane module replacement 34 MemoLog Memosens only 17 Memosens connecting cable 13 Memosens digital sensors 9 Memosens lab cable accessory 57 59 60 Memosens sensors connection 13 Memosens sensors product line 56 Menu icons 11 Messages Information menu 17 Micro USB port 13 N ...

Page 72: ...nt 9 Option 001 SOP 50 Option 001 SOP enabling 51 Option 002 Temp cal 50 Option 002 Temp cal enabling 55 Order numbers conductivity 58 Order numbers oxygen 60 Order numbers pH 57 ORP calibration 30 ORP configuration 21 Overview 8 Overview of error messages 49 Overview of status messages 47 Oxygen calibration Oxy 34 Oxygen configuration 26 Oxygen sensor amperometrical 60 Oxygen sensors product line...

Page 73: ...e correction Oxy configuration 26 Pre trigger limit value logger type 41 Printing 37 Printing a calibration record 38 Printing measured values 37 Product features 8 Product line 56 Product presentation 8 Protective cover 10 Q Quickstart guides 7 Quiver 8 R Rating plate 10 Real time clock 8 Rechargeable battery Li ion 12 Redox calibration 30 Registered trademarks 3 Replacement quiver 57 59 60 Repla...

Page 74: ...4 Sensor without temperature detector 36 Serial number of membrane optical oxygen 15 Serial number of sensor sensor information 15 Setup conductivity 23 Setup ORP 21 Setup oxygen 26 Setup pH 19 Shot logger type 40 SIP sensor information 15 Slope calibration oxygen 34 Softkeys 11 Software Paraly SW 112 46 SOP calibration Option 001 52 Specifications 61 Specific test report 7 Starting the data logge...

Page 75: ...ion 55 Temperature detectors accessory 56 58 60 Temperature manual adjustment 36 Temperature probe connection 13 Toggling the measured value display 36 Trademarks 3 U USB port battery 12 USB port micro 13 User management Option 001 SOP 53 V Value added features 9 Viewing the logger data 44 Z Zero calibration Cond 33 Zero calibration ISFET 31 Zero calibration Oxy 35 Index ...

Page 76: ...209 8MU KNE01 20160401 Software version 1 x 090453 Knick Elektronische Messgeräte GmbH Co KG Beuckestr 22 14163 Berlin Germany Phone 49 30 80191 0 Fax 49 30 80191 200 Email knick knick de Web www knick de ...