35
Oxy
DATA INPUT Calibration
(Calibration by entering known sensor values)
The calibration method is selected in the configuration menu.
Measurement
cal
CAL
DATA INPUT
xx blinks
nA
ZERO POINT
Use
to set the known value for the sensor
zero point.
cal
xxx blinks
nA
SLOPE
Use
to set the known value for the sensor
slope.
cal
Calibration is performed. The meter then automatically returns to measuring
mode.
Note:
To abort calibration, you can press
meas
at any time.
Oxygen Calibration