pH
Permissible calibration ranges Zero point
6 ... 8 pH
For ORP sensor:
ΔmV (offset)
-700 … 700 mV
With ISFET:
Operating point
(asymmetry)
-750 … 750 mV
Slope
Approx. 74 … 104 %
(possibly restricting notes from Sensoface)
Calibration timer
*
Interval 1 … 99 days, can be switched off
Sensoface
Provides information on the condition of the sensor
Evaluation of
Zero point/slope, response time, calibration interval
Specifications