![Knick COND 3400X-041 Short Description Download Page 48](http://html.mh-extra.com/html/knick/cond-3400x-041/cond-3400x-041_short-description_1980118048.webp)
48
COND 3400(X)-041
Menu Display
Diagnostics functions
Open diagnostics
From the measuring mode:
Press menu key to select menu.
Select diagnostics using arrow keys,
press enter to confirm.
The “Diagnostics” menu gives an
overview of all functions available.
Functions which have been set as
“Favorite” can be directly accessed
from the measuring mode.
Diagnostics Functions
General status information of the measuring system
Select menu: Diagnostics - Message list
Message list
Shows the currently activated warning
or failure messages in plain text.
Menu selection
Return to meas
0.120 mS/cm
25.6°C
Select:
[enter]
Lingua
Diagnostics
Return
25.6°C
Set favorite
0.120 mS/cm
Logbook
Message list
Device Description
FRONT module
Message list
Return
25.6°C
0.120 mS/cm
Warn
Modul BASE
8 messg.
Point of meas description
Warn Temp alarm HI
Saturation %Air Alarm HI
Fail Wrong ISM sensor
Warn Tolerance band
Warn Zero range
D062
P018
D013
P210
P030
Sensocheck
P092
Number of messages
When there are more than 7 messages,
a vertical scrollbar appears.
Scroll with the up/down arrow keys.
Module identifier
Specifies the module that has generated the
message.
Message identifier
Described in plain text