Diaphragm Pump N936
Installation and connection
Translation of Original operating and installation instruction, English, KNF 320276-320275 09/19
17
Connecting the pump
CAUTION
Risk of injury from mixing up suction and pressure
side
Mixing up the suction and pressure side can result
in breakage of connected components on the suc-
tion and pressure side.
à
Observe the marking of inlet and outlet on
the pump head.
1. Remove protective plugs from the hose connection threads.
2. Connect suction and pressure line (for mounting dimensions, see
Chapter
).
The pressure line serves to dissipate any condensate that may
form. A container that collects the condensate must therefore be in-
stalled in the pressure line.
3. Lay the suction and pressure line at a downward angle to prevent con-
densate from running into the pump.
Summary of Contents for N963
Page 29: ......
Page 30: ......
Page 31: ......