Chemically-resistant Laboratory pump N820.18 EX
Servicing
Translation of Original Operating Instruction, English, KNF
317428-317425 09/20
37
Component
Servicing interval
Pump
à
Inspect the pump periodi-
cally for external damage
or leakage.
à
Periodically check for no-
ticeable changes to
noises and vibrations.
Diaphragm and valve plates/
seals
à
At the latest, replace
when the pump flow rate
decreases.
Tab.19: Servicing schedule
11.2 Cleaning
NOTICE
During cleaning work, ensure that no
fluids enter the interior of the housing.