Set up, installation and connection
Chemically-resistant Laboratory pump N820.18
EX
28
Translation of Original Operating Instruction, English, KNF
317428-317425 09/20
9 Set up, installation and connection
Only connect the pumps in accordance with the operating pa-
rameters and conditions described in Chapter
.
Observe the safety notices (see Chapter
).
à
Before connecting, store the pump at the installation loca-
tion to allow it to reach the room temperature (no conden-
sation may form).
Cooling air
supply
WARNING
Danger of burning on hot surfaces
Hot surfaces could occur if the pump
overheats.
à
When installing the pump, make
sure that sufficient cooling air in-
feed and discharge is ensured.
Installation
location
à
Make sure that the installation location is dry and that the
pump is protected against rain, splash, gushing, and drip
water as well as from other contaminants.
à
Select a secure location (flat surface) for the pump.
à
Protect the pump from dust.
à
Protect the pump from vibration, impact and external dam-
age.
9.1 Connecting the pump
1. Remove the protective caps from the pneumatic connec-
tions of the pump.
Vacuum operation with silencer
1. Remove the hose connector on the pressure side with the
).
2. Screw the silencer or your system component to the pres-
sure side hand tight.
3. Connect the suction line.
Summary of Contents for LABOPORT N820.3FT.29.18G
Page 55: ...KNF weltweit Unsere lokalen KNF Partner finden Sie unter www knf com DE EN FR ES IT...
Page 109: ...KNF worldwide You can find our local KNF partners at www knf com...
Page 165: ...KNF dans le monde entier Vous trouverez nos partenaires KNF ici www knf com...
Page 222: ...KNF en el mundo Encontrar a nuestros socios KNF locales en www knf com DE EN FR ES IT...
Page 333: ...KNF wereldwijd Onze lokale KNF partners vindt u op www knf com...