Troubleshooting
AZURA® Detector UVD 2.1L Instructions, V6831
Message
Measure
Cannot proceed: lamps
are off.
1.
Switch the lamp on. If the error occurs
again, restart the device.
2.
If the lamp is off after restarting, repla-
ce the lamp.
Cannot read data from
FRAM
Restart the device.
Cannot read RTC
Restart the device.
Cannot write data on
FRAM
Delete the programs.
D2 lamp operation failed
1.
Restart the device.
2.
If the error occurs again, replace the
lamp.
D2 lamp does not start
1.
Restart the device.
2.
If the error occurs again, the lamp unit
has to be replaced. Contact the Custo-
mer Support.
Data acquisition active
No entries are possible.
Stop acquiring measurement data, then
you can make a new entry.
Filter move error
Restart the device.
GUI communication
failed
Restart the device.
Instrument remote
controlled
This entry is not executable.
Close and exit the software.
Invalid command
1.
Check the cable connections.
2.
Change the entry.
Invalid line number
Change the entry in the program line.
Invalid link
Reconfigure the link.
Invalid parameter(s)
Check the validity of the parameters.
Invalid time in time table
Correct the time entry.
Invalid index in time table
Change the entry in the program line.
Leak sensor not present
Restart the device.
Leak was detected
1.
Switch off the device.
2.
Remove the leak
3.
Switch on the device.
Link does not exist
Create a link.