9
Product information
AZURA® Detector UVD 2.1L Instructions, V6831
3.4 Meaning of the LEDs
There are three LEDs
1
and a standby switch
2
on the front of the
device. The figure shows the LED panel when the device is switched off.
The LEDs show different colors depending on the operating status.
Color
Operating status
Measures
left LED
Red
Error message
Check the system.
Shortly press the switch
to deactivate the error
message.
Green
Device is measuring.
-
center LED
Off
Device is not calibrated
or lamp is off.
Start calibration via soft-
ware or via rebooting the
device.
Switch on the lamp via
the software.
Flashes green Device is calibrating.
Wait until the device is
ready.
Green
Device is ready for
operation.
-
right LED
Green
Device is active.
-
Blue
Standby
Exit standby by pressing
the standby switch.
To start the standby, keep the switch pressed for 5 seconds.
Note:
Malfunctioning system after repeated standby possible. After
repeatedly using the standby, switch off the power switch and back on
again, to reset the data storage.
1
2
Standby