AZURA® Detector UVD 2.1L Instructions, V6831
7.3 Error messages
If other error messages are displayed besides those listed below, please
restart the device.
Inform the Customer Support of the manufacturer in case the error mes-
sage displays repeatedly.
Message
Measure
Calibration failed
1.
Restart the device.
2.
Check whether lamps, motor and filter
are functioning correctly.
3.
Restart calibration on the device or in
the chromatography software.
Calibration failed - Filter
position error
The filter wheel moves incorrectly.
Restart the device.
Calibration failed - Low
light
Both reference channel and signal channel
do not have enough light for the calibrat-
ion due to, for instance, a defective lamp.
Restart the device.
Calibration failed: Signal
path open
Restart the device.
Calibration failed: Too
much light
Restart the device.
Cannot delete active
program/link
1.
Pause link.
2.
Delete program.
Cannot edit program
from the running link
1.
Pause link.
2.
Edit data using chromatography
software.
Cannot initialize LAN
Check cables and connections in local
area network.
Cannot operate an unca-
librated instrument
1.
Restart the device.
2.
Wait until calibration is completed.
Cannot operate with an
empty link
Create a link.
Cannot proceed: D2
lamp heating
Restart the device. If the error occurs
again, replace the lamp.
Cannot proceed: D2
lamp is OFF!
1.
Switch the lamp on. If the error occurs
again, restart the device.
2.
If the lamp is off after restarting, repla-
ce the lamp.