CONFIGURATION MODE
This document contains confidential information. Dissemination, publication, copying, or duplication of this document or software described herein without prior written authorization is strictly prohibited.
kitchenbrains.com
Page 8 of 16
© 2020 Kitchen Brains
VIEW SETTINGS:
Displays the settings for the following parameters currently
programmed in the 75LCT-for a quick glance. Each parameter
will have its own setup screen, as well as input limits (as de-
fined in table 1).
Date –
Date in MM/DD/YYYY for US market; DD/MM/YYYY
for non-US markets
Scale –
C or F
Calibration Offset –
Offsets displayed temperature by preset
amount
High Limit –
High temperature threshold for sounding alarm
Low Limit -
Low temperature threshold for sounding alarm
Temp Delay –
Time delay in minutes, after which alarm
condition is triggered for temperature exceeding the high
or low limit thresholds
Door Ajar Delay –
Time delay in minutes, after which alarm
condition is triggered for door open/Ajar
Lights Off Delay –
Lights off timer, in minutes
Lights Operation –
Lights operation mode
Ring Back Delay –
Delay for timed repeat audible alarms during
an ongoing temperature alarm or power failure alarm condition
Alarm Relay Mode –
Set mode of operation for relay
Auto Panic –
Set auto panic mode
PROGRAM:
Allows user to configure/change default parameters. Table
2 lists the default settings as well as the selection range for
each of the parameters.
Table 2. Program Menu Options
MENU ITEM SELECTION (RANGE) DEFAULT
Scale
F or C
F
Calibration
Offset
2
-5 to +5
0
High Limit
-40F to 125F or
-40C to 52C (1º interval)*
42F/6C cooler
10F/-12C Freezer*
Low Limit
-40F to 125F or
-40C to 52C (1º interval)*
30F/1C cooler
-20F/-29C Freezer*
Tem Delay
0 to 150 mins
(1 m interval)*
60
Door Adjar
Delay
0 to 60 mins
(1 m interval)*
20
3
Lights Off
Delay
0 to 150 mins
(1 m interval)*
14
4
Lights
Operation
Timed, Untimed
Timed
Ring Back
Delay
0 to 150 mins
(1 m interval)*
30
Alarm Relay
Mode
Unlatched - Latched
Unlatched
AutoPanic
On
6
- Off
On
Sleep Mode:
The 75LCT ships from the factory in Sleep Mode.
The rechargeable battery is connected physically but is not
connected electrically until AC power is applied.
2 Calibration Offset
shall not change Temperature Setpoints
3 Value of 0 =
feature deactivated/no delay
4 Value of 0 =
feature deactivated/no delay
5 In Latched mode,
temp alarm condition, alternate display between “High/Low Temp”
and Temp value in Red color
6 AutoPanic =
On & Lights Operation =Timed allowed only if MC-1 and either one or both
IP-1 and MD-1/MD-2 installed
75LCT MULTI-MONITOR |
OPERATING MANUAL