Tab. 11: Error messages of the unit
Display
Buzzer
Description
Measure
Unit key
required
3
Sensor X:
n
Error or short circuit in the rel-
evant sensor.
n
The cryostat is running in the
emergency program.
n
Contact the service depart-
ment.
-
3
Sensor X:
n
Error or break in the relevant
sensor.
n
Cryostat is running in the
emergency program.
n
Contact the service depart-
ment.
-
3
Fan X:
n
Speed of the relevant fan too
low.
n
The temperature of the chilled
goods can fluctuate.
n
Transfer chilled goods to new
location.
n
Contact the service depart-
ment.
-
3
Fan X:
n
Speed of the relevant fan too
high.
n
The temperature of the chilled
goods can fluctuate.
n
Transfer chilled goods to new
location.
n
Contact the service depart-
ment.
-
3
Fan:
n
Fan does not reach required
minimum speed after an unit
restart.
n
The temperature of the chilled
goods can fluctuate.
n
Transfer chilled goods to new
location.
n
Contact the service depart-
ment.
-
3
Fan:
n
Difference between the
speeds of the fans is too
great.
n
The temperature of the chilled
goods can fluctuate.
n
Transfer chilled goods to new
location.
n
Contact the service depart-
ment.
-
3
Relay X:
n
Defect in the relevant relay.
n
The temperature of the chilled
goods can fluctuate.
n
Transfer chilled goods to new
location.
n
Contact the service depart-
ment.
-
3
Relay X:
n
Defect in the relevant relay.
n
The temperature of the chilled
goods can fluctuate.
n
Transfer chilled goods to new
location.
n
Contact the service depart-
ment.
-
Status displays and error messages
Error messages
26.05.2021
Refrigerator/freezer BL series
70