3.6 EMC and requirements of the electrical standards
WARNING!
Danger due to failure to comply with require-
ments for electromagnetic compatibility!
Electrical medical equipment is subject to specific
requirements for electromagnetic compatibility
(EMC). Failure to comply with the safety require-
ments leads to a risk of property damage and
injury. In addition, failure of the medical equipment
may result in damage to or destruction of the
chilled goods.
Make sure that electrical medical equipment is
installed and commissioned in accordance with the
following information:
– It is prohibited to use any accessories other
than those described and sold by the manufac-
turer. Spare parts that are not genuine Kirsch
spare parts may increase electromagnetic
emitted interference or reduce the electromag-
netic immunity of the unit.
– Portable and mobile HF equipment (e.g. mobile
phones) can have an impact on electrical med-
ical equipment. The recommended safety dis-
tances must be maintained.
– The unit must not be set up next to, above or
below another unit. If this is unavoidable, check
the unit before use under operating conditions
to verify that it functions correctly.
NOTICE!
This unit is not a life-saving or life-sustaining unit.
Requirements for electromagnetic
compatibility (EMC)
Safety
EMC and requirements of the electrical standards
26.05.2021
Refrigerator/freezer BL series
28