Kipp & Zonen METEON Instruction Manual Download Page 8

  

Wizard 

Alternatively, the settings are generated by software through the Wizard. Depending on the 
sensor sensitivity, the Wizard will select the optimal input stage amplification for the 
METEON, and store appropriate values in the multiplier and divider fields. The position of the 
decimal dot in the display is also set. After the wizard is used, the METEON only has to be 
programmed to make it suitable for use with the specified sensor. 
 
The following sensors are supported: 
 

Sensor Full 

scale 

Sensitivity 

CMP 3 

2000 W/m2 

5 .. 20 µV / W/m2 

CMP 6 

2000 W/m2 

5 .. 16 µV / W/m2 

CMP 11 

4000 W/m2 

7 .. 14 µV / W/m2 

CMP 21 

4000 W/m2 

7 .. 14 µV / W/m2 

CMP 22 

4000 W/m2 

7 .. 14 µV / W/m2 

CM 4 

4000 W/m2 

7 µV / W/m2 

SP Lite 

2000 W/m2 

60 .. 100 µV / W/m2 

SP Lite * 

4000 W/m2 

9.5 .. 10.5 µV / W/m2 

PAR Lite 

10,000 µmol/m2s 

4 .. 6 µV / µE/m2s 

NR Lite 

±500 W/m2 

10 µV / W/m2 

CUV 4 

200 W/m2 

1 mV / W/m2 

 
 
 
Use the wizard to select the sensor model that is to be used, and then enter the sensitivity of 
the sensor. This is only required when the sensor is changed or when the METEON is bought 
without a sensor. Otherwise the sensitivity of your sensor is already in the METEON. 
 

 

Data logger 

The PC software is used to enter the data logger setup, and to transfer the setup to the 
METEON. 
 
 
When the METEON is programmed with a new data logger setup, the logger memory in the 
METEON is erased, and then the new settings are used to start a new logging sequence. The 
logging only stops when the logger memory is full, when the batteries are depleted, or when  
the METEON is switched off. The PC software can be used to check the logging activities, 
and the METEON also shows some display characters to indicate what it is doing. 
 
Note: 
When the METEON is switched off, all logged values and settings remain in the (flash) 
memory, even when the batteries are removed. 
 
 
 
 

Summary of Contents for METEON

Page 1: ...Instruction Manual Irradiance Meter METEON...

Page 2: ...our sole responsibility that the products Type Handheld Readout unit Name METEON to which this declaration relates are in conformity with the following standards Imissions EN 50082 1 Group standard Em...

Page 3: ...mV 6 250 1 V Basic accuracy after factory calibration 0 1 Temperature sensitivity 0 5 over the full temperature range Offset drift 3 V over the full temperature range Operation temperature range 10 4...

Page 4: ...f the METEON to avoid damage to the case Extract the battery holder from the compartment and install two fresh batteries Notice the polarization Replace the bottom cover and connect the sensor The MET...

Page 5: ...es bundled with a sensor the sensitivity is already programmed into the METEON In that case the optional interface cable and the software give you access to the data logging functions only Main window...

Page 6: ...the PC does not detect the METEON then check if the METEON is switched on and check the batteries Once the METEON is detected the caption of the window will show the serial number of the METEON that w...

Page 7: ...ftware In the box that says Input amplifier you can select the amplification that corresponds to the required full scale input 2 Next to the amplification factor you can read how many internal counts...

Page 8: ...500 W m2 10 V W m2 CUV 4 200 W m2 1 mV W m2 Use the wizard to select the sensor model that is to be used and then enter the sensitivity of the sensor This is only required when the sensor is changed o...

Page 9: ...one the start of the first logging interval until a specified date and time This date and time at which the first interval will start can be entered in the PC software When the logger is busy counting...

Page 10: ...ew the logger status while the logger is active Reading the status will not interfere with the logging process The status will show what the logger is doing how much time remains in the current interv...

Page 11: ...e bottom right corner of the area of interest When the mouse is released the chart will zoom in on the selected area Drag the mouse from bottom right to top left to zoom out again Save the data in a l...

Page 12: ...hrough USB BAT indicates battery low replace with 2 AA Alkaline types m writing data in METEON memory l logging started 0 5 s interval waiting for initial delay to pass 1 s interval The actual reading...

Page 13: ...nen B V Delftechpark 36 2628 XH Delft P O Box 507 2600 AM Delft The Netherlands T 31 0 15 2755 210 info kippzonen com www kippzonen com Kipp Zonen Asia Pacific Pte Ltd 10 Ubi Crescent Lobby E 02 93 Ub...

Reviews: