Introduction 

 
The Kipp & Zonen METEON is a high-performance sensor readout unit with integrated data 
logger. 
 
The sensor sensitivity is programmed into the METEON such that the readout will show 
calibrated engineering units, rather than millvolts or microvolts. 
 
Use of the data logger functions and programming the sensor sensitivity is done with the USB 
interface which is standard supplied to connect the METEON to a PC. 
 
 

Specifications 

 

 

A/D conversion 

16 bits 

Input range (programmable) 

 

Full scale voltage 

Internal counts 

Internal resolution 

+/- 200 mV  

20,000  10µV 

+/- 100 mV 

10,000  10µV 

+/- 50 mV 

5,000  10µV 

+/- 25 mV 

25,000  1µV 

+/- 12.5 mV 

12,500  1µV 

+/- 6.25 mV 

6,250  1µV 

Basic accuracy after factory calibration 

0.1% 

Temperature sensitivity 

< 0.5% over the full temperature range  

Offset drift 

< 3 µV over the full temperature range 

Operation temperature range 

-10 .. +40 °C 

Storage temperature range 

-20 .. +60 °C 

Relative humidity 

< 95% RH non-condensing 

Input circuit 

> 2 M

 // 680 nF 

Input low-pass filter 

5 Hz 

Input connectors 

2 x 4mm banana plug 

Display 

LCD 4 digits with sign 

Decimal dot position (programmable) 

8888 or 888.8 or 88.88 or 8.888 

Display update 

At least 1 per second 

Computer interface 

USB 1.1 / 2.0 low speed 

Supply current 

< 2 mA 

Batteries 

2 AA (penlight) alkaline 

Battery life (alkaline) 

> 1200 hours (50 days) continuous use 

Data logger memory 

3518 samples 

Data logger storage interval 

2 .. 65535 seconds 

Logged information 

Min, max and average over log interval 

Software compatibility 

Windows 2000, Windows XP 

Summary of Contents for METEON

Page 1: ...Instruction Manual Irradiance Meter METEON...

Page 2: ...our sole responsibility that the products Type Handheld Readout unit Name METEON to which this declaration relates are in conformity with the following standards Imissions EN 50082 1 Group standard Em...

Page 3: ...mV 6 250 1 V Basic accuracy after factory calibration 0 1 Temperature sensitivity 0 5 over the full temperature range Offset drift 3 V over the full temperature range Operation temperature range 10 4...

Page 4: ...f the METEON to avoid damage to the case Extract the battery holder from the compartment and install two fresh batteries Notice the polarization Replace the bottom cover and connect the sensor The MET...

Page 5: ...es bundled with a sensor the sensitivity is already programmed into the METEON In that case the optional interface cable and the software give you access to the data logging functions only Main window...

Page 6: ...the PC does not detect the METEON then check if the METEON is switched on and check the batteries Once the METEON is detected the caption of the window will show the serial number of the METEON that w...

Page 7: ...ftware In the box that says Input amplifier you can select the amplification that corresponds to the required full scale input 2 Next to the amplification factor you can read how many internal counts...

Page 8: ...500 W m2 10 V W m2 CUV 4 200 W m2 1 mV W m2 Use the wizard to select the sensor model that is to be used and then enter the sensitivity of the sensor This is only required when the sensor is changed o...

Page 9: ...one the start of the first logging interval until a specified date and time This date and time at which the first interval will start can be entered in the PC software When the logger is busy counting...

Page 10: ...ew the logger status while the logger is active Reading the status will not interfere with the logging process The status will show what the logger is doing how much time remains in the current interv...

Page 11: ...e bottom right corner of the area of interest When the mouse is released the chart will zoom in on the selected area Drag the mouse from bottom right to top left to zoom out again Save the data in a l...

Page 12: ...hrough USB BAT indicates battery low replace with 2 AA Alkaline types m writing data in METEON memory l logging started 0 5 s interval waiting for initial delay to pass 1 s interval The actual reading...

Page 13: ...nen B V Delftechpark 36 2628 XH Delft P O Box 507 2600 AM Delft The Netherlands T 31 0 15 2755 210 info kippzonen com www kippzonen com Kipp Zonen Asia Pacific Pte Ltd 10 Ubi Crescent Lobby E 02 93 Ub...

Reviews: