![Kingfisher KI 7340 Series Operation & Maintenance Manual Download Page 39](http://html1.mh-extra.com/html/kingfisher/ki-7340-series/ki-7340-series_operation-and-maintenance-manual_1970295039.webp)
KI 7340/7740 UM-12
37
SPECIFICATIONS: POWER METER
KI7340C Series Power Meter section
Detector
Type
Response
nm
Damage Level,
dBm
Calibration wavelengths,
nm
Power
Range
dBm
Autotest
Sensitivity
dBm
Mid range
Linearity¹
dBm
Calibration
Accuracy
2
%
Polarisation
Sensitivity
dB
Total
Uncertainty
3
dB
InGaAs
600 ~ 1700
+15
780, 820, 850,
980,
1270, 1290, 1300, 1310, 1330, 1350,
1370, 1390, 1410, 1430, 1450, 1470, 1490,
1510, 1530, 1550, 1570, 1590, 1610, 1625
, 1650
+5 ~ -60
+5 ~ -70
-45
-50
0.02
1 %
(0.06 dB)
<0.05
0.3
H3B
(InGaAs)
800 ~ 1700
+30
4
820, 850, 980,
1270,1290,1300, 1310, 1550, 1270, 1290, 1330, 1350,
1370, 1390, 1410, 1430, 1450, 1470,
1490, 1510, 1530, 1570, 1590, 1610, 1625,
1650
+27 ~ -40
+27 ~ -50
-25
-30
0.02
1 %
(0.06 dB)
<0.05
0.35
H5
(InGaAs)
800 ~ 1700
+25
820,850,980,
1270, 1290, 1300, 1310,1330,1350,1370, 1390, 1410,
1430, 1450, 1470, 1490, 1510, 1530, 1570, 1590, 1610,
1625
,1650
+15 ~ -50
+15 ~ -60
-35
-40
0.02
1 %
(0.06 dB)
<0.05
0.3
Si
350 ~ 1100
+15
635, 650, 660, 780, 850, 980
+0 ~ -70
-50
0.02
1 %
(0.06 dB)
<0.05
0.3
typical
typical
Typical
max
KI7740C Series Power Meter section
Detector
type
Response
wavelengths,
nm
Damage
level,
dBm
Calibration wavelengths,
nm
Power
range,
dBm
Autotest
Sensitivity,
dBm
Mid range
linearity¹
dBm
Calibration
accuracy
2
%
Polarisation
sensitivity,
dB
Total
uncertainty
3
dB
Wavelength
sensitivity
± 30 nm
5
dB
InGaAs
600 ~ 1700
+15
780, 820, 850, 980,
1270, 1290, 1300, 1310, 1330, 1350, 1370, 1390,
1410, 1430, 1450, 1470, 1490, 1510, 1530, 1550, 1570,
1590, 1610, 1625
, 1650
+5 ~ -60
+5 ~ -70
-45
-50
0.02
1 %
(0.06 dB)
<0.05
0.3
0.03
H3B
(InGaAs)
800 ~ 1700
+30
4
820, 850, 980,
1270, 1290, 1300, 1310, 1330, 1350, 1370, 1390,
1410, 1430, 1450, 1470, 1490, 1510, 1530, 1550, 1570,
1590, 1610, 1625,
1650
+27 ~ -40
+27 ~ -50
-25
-30
0.02
1%
(0.06 dB)
<0.05
0.3
0.03
H5
(InGaAs)
800 ~ 1700
+25
820, 850, 980,
1270, 1290, 1300, 1310, 1330, 1350, 1370, 1390,
1410, 1430, 1450, 1470, 1490, 1510, 1530, 1550, 1570,
1590, 1610, 1625,
1650
+15 ~ -50
+15 ~ -60
-35
-40
0.02
1 %
(0.06 dB)
<0.05
0.3
0.03
Si
350 ~ 1100
+5
635, 650, 660, 780, 850, 980
+0 ~ -70
-50
0.02
1 %
(0.06 dB)
<0.05
0.3
0.03
typical
typical
typical
max
typical
Note 1:
Mid-range linearity @ 1550 nm for InGaAs & Ge, or 850 nm for Si. Non-coherent light, with APC connector. Excludes top 5 dB and bottom 10 dB of range.
Note 2:
Calibration condition: non-coherent light, -35
5 dBm, 21
3 °C,
1 nm, 10
3 nm FWHM, PC ceramic connector, 100
m fiber
Note 3:
Includes contributions of: varying optical connector types, calibration uncertainty, linearity over temperature & range, and fiber
core diameter up to 200 μm.
Note 4:
H3B can sustain the damage level for 2 minutes
Note 5:
At calibration wavelengths in bold type