ACCURACY CONSIDERATIONS
Use of ORL Zero function
This function is very useful since it extends the practical linear
measurement range by about 14 dB.
For typical mid-range measurements, it has a negligible effect on
measurement noise. For low level measurements, it can cause up to
0.5 dB noise. If this happens, the solution is to reduce the residual light
level further, and re-set the zero function.