MEASURING TWO WAY LOSS AND ORL IN AUTOTEST
Pass / Fail display mode
The Two Way Pass / Fail display mode is convenient for quick
acceptance testing. It allows performing testing on all available
wavelengths.
To enter Pass / Fail mode, press [P/F]. To exit, press [Exit].
To set the loss threshold, press [Power Meter], [P/F] and [-/+] to
adjust, then [Set] to save.
Alternatively, the value defaults to 12.00 dB by pressing [
] on the
hidden keypad and then [Default]. Press [Set] to save.
To set the ORL threshold, press [Return Loss], [P/F] and [-/+] to adjust,
then [Set] to save.
Alternatively, the value defaults to -25.00 dB RL by pressing [
] on
the hidden keypad and then [Default]. Press [Set] to save.
: