KI 7010B UM-11
21
PERFORMANCE VERIFICATION TEST
All tests can be performed without opening the instrument.
The test procedures described in this section are for performance
verification of a KI7010B Automated Variable Optical Attenuator.
Due to large number of possible instrument configurations, it is not
possible to give detailed test procedures for all options in this manual,
so some parameters may need adjusting to the appropriate
specifications.
Required Equipment:
this is the required equipment for the
performance test listed. Any equipment that satisfies the critical
specifications of the instruments given in the table may be substituted
for the recommended models.
Test Record:
results of the performance test may be tabulated on a
photocopy of the Test Record provided at the end of the test
procedure. It is recommended that you fill out the Test Record and
refer to it while doing the test.
Test Failure:
if the instrument fails any performance test, return it to
the nearest Sales/Service office for repair.
Instrument Specification:
specifications are the performance
characteristics of the instrument that are certified, and are the limits
against which the equipment under test can be tested.
Any changes in the specifications due to manufacturing changes,
design or traceability to NATA will be covered in a user manual change
supplement or revised manual. Such specifications supersede any
that were previously published.
General Instructions
Perform each step in the order given, using the corresponding test
equipment. Use Tables 1 ~ 3 to record test details.
The SMF/MMF test lead fiber type and PC/APC connector polish must
be matched to the instrument type.
Ensure that all optical connections are dry and clean
. DO NOT USE
INDEX MATCHING OIL.
For cleaning, please refer to the cleaning
instructions given in the section
‘
Optical Connector
’.
Check that all patch leads are fixed to the table to avoid movements
during measurements.
Make sure that the ambient conditions are in the following ranges:
Temperature:
21 ± 3 ºC
Relative humidity:
45 to 75 %
Summary of Contents for KI 7010B Series
Page 1: ...OPERATION MAINTENANCE GUIDE KI 7010B Series Automated Variable Optical Attenuator ...
Page 5: ...KI 7010B UM 11 3 KI7000B SERIES AUTOMATED VARIABLE OPTICAL ATTENUATOR ...
Page 32: ...KI 7010B UM 11 30 ...
Page 34: ...KI 7010B UM 11 32 NOTES ...
Page 35: ...KI 7010B UM 11 33 NOTES ...
Page 36: ...KI 7010B UM 11 34 NOTES ...