Guidance and manufacture’s declaration – electromagnetic immunity
Immunity test
80 MHz to 2.5
GHz
3V/m
3V/m
80 MHz to 800 MHz
800 MHz to 2.5 GHz
Portable and mobile RF communications
equipment should be used no closer to any
part of the Compressor Nebuliser,
including cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of the
transmitter.
Where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
distance in metres (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
Radiated RF
IEC 61000-4-3
IEC 60601 test Complianc
level
level
Electromagnetic environment - guidance
Recommended separation distance
The Compressor Nebuliser is intended for use in the electromagnetic environment specified
below. The customer or the user of Compressor Nebuliser should assure that it is used
in such an environment.
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
survey, should be less than the compliance
level in each frequency range.
b
Interference may occur in the vicinity of
equipment marked with the following
symbol:
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the
Compressor Nebuliser
is used exceeds the applicable RF compliance level above, the
Compressor Nebuliser
should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as reorienting or relocating the
Compressor Nebuliser.
a
b
Over the frequency range 150 KHz to 80 MHz, field strengths should be less than 3V/m.
14. EMC Table -1