To test the connection of the wire shields to the EpiSensor-end connector
shell:
1. At the recorder end, have another technician take the recorder
connector and use a test-lead jumper to connect pin L to the
recorder-end connector shell.
2. The DVM at the EpiSensor-end should show that pin L is now
connected to the recorder-end connector shell and to pin U. The
DVM should measure < 1.5
%
per 100' of cable. If it does, have the
technician remove the test jumper at the recorder end.
3. Disconnect the DVM lead from connector pin L and confirm that
your connector shell and pin U are infinite with respect to all the
other connector pins.
With the final cable testing completed, disconnect all test equipment. The
fully assembled cable is ready to connect to the recorder and the EpiSensor.