background image

7

DIGITAL AUTO RANGING MULTIMETER

Diode and Continuity

Range

Introduction

Remark

The approximate forward voltage drop 

will be displayed

Open circuit voltage: approx 1.5V

1. 

The built-in buzzer will sound if 

the resistance is less than about 
30

Ω. 

Open circuit voltage: approx 0.5V

Overload Protection: 250V DC/AC rms

For Continuity Test: When the resistance is betwene 30Ω and 100Ω, the buzzer may sound or may not sound. When the

resistance is more than 100Ω, the buzzer wont sound. 

Capacitance

Range

Resolution

Accuracy

40

nF

10pF

± (8% of rdg + 10dgts)

400nF

1

00pF

± (5% of rdg + 5dgts)

4uF

1nF

40uF

10nF

400uF

100nF

4000uF

1uF

± (8% of rdg + 10dgts)

Overload Protection: F0.5A/600V fuse

Open Circuit Voltage: approx. 0.5V

Frequency (Auto Ranging)

Range

Accuracy

10/100/1K/10K/100K/1M/10Mhz

± (0.1% + 5)

Overload Protection: 250V DC/AC rms

8) Operation

Measuring Voltage

WARNING! 

 To avoid electrical shock and/or damage to the instrument, do not attempt to take any voltage measurements    

 which may exceed 600V DC or 600V AC rms. Do not apply more than 600V DC or 600V AC rms between the common     

  terminals and the earth ground.

1. 

Connect the BLACK test lead (10) to the “COM” jack (6) and the RED lead (10) to the “VΩ” jack (7).

2. 

Set the function switch (3) to V~ or V

 range. Select auto range or manual range with the “RANGE” button (2). 

3. 

In manual range, if the voltage magnitude to be measured is unknown beforehand, select the highest range.

4. 

Connect the test leads (10) across the source or load to be measured.

5. 

Read LCD display (1). The polarity of the RED lead (10) connection will be indicated when making a DC measurement.

Note:

a)       In small range, the meter may display an unstable reading when the test leads have not been connected to the  load to be  
          measured. It is normal and will not affect the measurements.
b)       In manual range mode, when the meter shows the over range symbol “OL”, a higher range must be selected.
c)       To avoid damage to the meter, dont measure a voltage which exceeds 600Vdc (for DC voltage measurement)  or  

            

          600Vac (for AC voltage measurement). 

Summary of Contents for K8315

Page 1: ...AUTO RANGING DIGITAL MULTIMETER K8315 ED1 May 17 LARGE DIGITAL DISPLAY 12 MONTH WARRANTY AC DC VOLTAGE CURRENT MEASUREMENT CAT II SAFETY RATING CAT III TEST LEAD SAFETY RATING...

Page 2: ...ct or 9 DATA H Button Hold 10 Test Leads CAT III 11 Multi Function Capacitance Transistor Adaptor Model No K8315 Analog Display 42 segments updates 12 5 times sec Digital Display 3999 Counts updates 2...

Page 3: ...before using the Meter e Verify the Meter s operation by measuring a known voltage before and after using it Do not use the Meter if it operates abnormally Protection may be impaired If in doubt have...

Page 4: ...tion Wear approved safety eye protection 3 Service a Tool service must be performed only by qualified repair personnel b When servicing a tool use only identical replacement parts Use only authorized...

Page 5: ...nd Low Battery Fuse Diode Continuity Test Conforms to European Union Directive Double Insulated 6 Multimeters Function Table Model DCV ACV DCA ACA hFE CAP Hz K8315 8 Multimeter Product Specifications...

Page 6: ...ge hFE Test Current Test Voltage PNP NPN 0 1000 lb 2 A Vce 1V DC Current Range Resolution Accuracy 400 A 0 1 A 0 8 of rdg 5dgts 4000 A 1 A 40mA 10 A 400mA 100 A 4A 1mA 1 5 of rdg 3dgts 10A 10mA Cautio...

Page 7: ...u must wait 15 minutes before reusing the multimeter Overload Protection mA jack F0 5A 600V Fuse 10A jack F10A 600V Fuse Max Input Current mA jack 400mA 10A jack 10A Voltage Drop 400 A 40mA and 4A ran...

Page 8: ...nt do not attempt to take any voltage measurements which may exceed 600V DC or 600V AC rms Do not apply more than 600V DC or 600V AC rms between the common terminals and the earth ground 1 Connect the...

Page 9: ...high voltage capacitors before measuring resistance 1 Connect the BLACK test lead 10 to the COM jack 6 and RED lead 10 to the V jack 7 Note The polarity of the RED test lead 10 is positive 1 Set the...

Page 10: ...he COM Jack 6 and the mA jack 5 Ensure the writing on the Multi Function Capacitance Transistor 11 is correctly orientated Identify whether the transistor is NPN or PNP type and locate Emitter base an...

Page 11: ...Tel 1300 657 528 The applicable warranty period 12 months commences on the date that the product is purchased If this product has materials or workmanship defects other than defects caused by abnorma...

Page 12: ...11 DIGITAL AUTO RANGING MULTIMETER Notes...

Page 13: ...www kincrome com au...

Reviews: