Maintenance
This chapter describes the maintenance and inspection of the tester. To maintain the tester’s
original performance for as long as possible, conduct periodic inspection and maintenance.
Inspection
AC power cord
Confirm that the covering is free of breaks and cracks, that the plug is free of looseness and
cracks, and that the cord does not contain a broken wire.
High-voltage test leadwires
Confirm that the covering is free of breaks and cracks and that there is no break in the
leads.
To purchase accessories, contact your Kikusui distributor/agent.
Maintenance
Backup battery
The tester uses a lithium battery for memory backup.
If the battery is exhausted, the memory will not retain stored test conditions or other data.
Although the service life depends on the usage environment, we generally recommend that
the battery be replaced every three years. We also recommend that regular internal
inspections and cleaning be conducted at the same intervals.
Calibration
Regular calibration is required to maintain the accuracy of measuring instruments over
extended periods. For calibration, contact your Kikusui distributor/agent.
WARNING
Any break or crack in the covering greatly increases the risk of electric shock or fire. If
you find any breaks or cracks, immediately discontinue use of the tester.
WARNING
The tester generates voltages as high as 1000 Vdc. Never attempt to maintain the
Tester including overhauls for yourself, as such tasks entail great danger. For such
service, contact your Kikusui distributor/agent.
WARNING
The tester generates voltages as high as 1000 Vdc. Never attempt to calibrate the Tester
for yourself, as such tasks entail great danger. For such service, contact your Kikusui
distributor/agent.