TOS7200
41
REMOTE Terminal
Performing remote control
1
Turn OFF the POWER switch.
2
Connect the REMOTE terminal and the remote-control box using the
dedicated connection cable (5-pin DIN cable) via the DIN-mini DIN
adaptor.
3
Turn ON the POWER switch.
This allows the input of a start signal from the remote control box and disables the
START switch on the front panel. Note that stop operations may be performed
both by using the STOP switch on the front panel and by transmitting a stop
signal from the remote control box. For more information, see the operation
manual for the optional remote control box.
4
To restore control from the front panel, turn OFF the POWER switch.
5
Remove the dedicated connection cable (6-pin DIN cable) from the
REMOTE terminal on the front panel.
6
Turn ON the POWER switch.
This enables the START switch on the front panel.
• When a START signal at the SIGNAL I/O connector is enabled, the START signal has priority
over a START signal at the REMOTE terminal. A STOP signal through the SIGNAL I/O
connector and that through the REMOTE terminal are accepted equally.
• Connecting or disconnecting a remote control box to or from the REMOTE terminal with the
POWER switch turned ON interrupts high-voltage output.
• When FAIL MODE in the system setting
has been activated, a FAIL judgment cannot
be canceled by the input of a stop signal via the REMOTE terminal. In such cases, use the
STOP switch on the front panel to cancel the judgment.