42
TOS7200
SIGNAL I/O Connector
The SIGNAL I/O connector is a D-SUB 25-pin connector on the rear panel.
This connector is used to start and stop tests, recall panel memory entries, and monitor the
status of the TOS7200.
•
Connector on the tester side
XM3B-2522 D-sub 25-pin female connector (socket) ;
manufactured by OMRON Corporation, Screw M2.6 x 0.45
•
Complies connector (plug)
D-sub 25-pin male (with fix screw M2.6)
To avoid malfunctions caused by noise, use shielded-type D-sub 25-pin connectors and a
cable that is 2.5 m or less in length.
For information about how to obtain replacement parts, contact your Kikusui agent or
distributor.
For information about how to use these components, see the OMRON Corporation catalogs.
Wire and tool that are necessary to make the connection
WARNING
Possible electric shock. Ensure that all devices are off before you connect or disconnect
cables between them.
CAUTION
Possible damage to the internal circuitry.
• Keep the signal wire at least 50 cm away from the high-voltage test lead and the DUT.
• Do not short the output voltage circuit.
• If you use the TOS7200 with incomplete connections, burn outs caused by heating may
occur when the output is turned on.
• Do not touch the contacts or attach insulators to them. Doing so may reduce the quality of
the contacts or cause other problems.
RS232C
SIGNAL I/O
SIGNAL I/O connector
Wire
Single wire:
Φ
0.32 mm (AWG28) to 0.65 mm (AWG22) in diameter
Twisted wire:
Φ
0.32 mm
2
(AWG22) to 0.08 mm
2
(AWG28)
Wire stripper
A wire stripper that matches the wires listed above