78
PFX2532
PFX2532 Electric Specifications
Unless specified otherwise, the specifications are for the following settings and conditions.
• The warm-up time is 30 minutes.
• TYP (typical) values do not guarantee the performance.
• reading: Indicates the readout value.
• set: Indicates the setting value.
• rating: Indicates the rated.
• Static: General term to indicate CC charge, CC - CV charge, CC discharge, CC - CV discharge, CP
discharge, and CP - CV discharge
• Pattern: General term to indicate pattern charge / discharge and I - V characteristics charge /
discharge
Rated Output
Number of outputs
1 ch
Charge current range
1
1
The range varies depending on factors such as the connected DC power supplies and electronic loads,
the wiring configuration of the system, and the charge/discharge operation.
0.000 A to 200.000 A
Charge voltage range
1
60 V range
0.000 V to 60.000 V
6 V range
0.000 V to 6.000 V
Discharge current range
1
0.000 A to 200.000 A
Discharge voltage
range
1 2
2
The minimum voltage that can be discharged varies depending on factors such as the connected
electronic load models and the wiring configuration.
60 V range
0.000 V to 60.000 V
6 V range
0.000 V to 6.000 V