Connecting the Power Cord
2
C
o
nn
ec
ting
an
d C
o
nfigurin
g Devic
e
s
■
Connection procedure
A terminal block cover is attached to the AC INPUT terminal block when the
product is shipped from the factory. This cover prevents the terminals from
being touched unintentionally. If it is damaged or lost, contact your Kikusui
agent or distributor.
1
Check that the AC power line meets the nominal input rating of the PFX2532.
The PFX2532 can receive a nominal line voltage in the range of 100 Vac to 240 Vac at
50 Hz or 60 Hz.
2
Check that the POWER switch is turned off.
3
Remove the terminal block cover from the AC INPUT terminal block.
4
Connect the power cord to the L, N, and
(GND) terminals of the AC INPUT
terminal block on the rear panel.
5
Attach the terminal block cover that you removed in step 3 .
Use the left hole to attach it.
6
Attach a appropriate solderless terminal to the switchboard end of the power
cord.
The included power cord comes with a solderless terminal on the PFX2532 end, but
no termination is provided on the switchboard end. Attach a solderless terminal that is
appropriate for the screw on the switchboard that you are connecting to. Have a
qualified engineer perform the work.
7
Turn off the switchboard’s circuit breaker.
8
Connect the L, N, and
(GND) wires of the power cord to the matching
terminals on the switchboard.
[Q5-000-159]
Front view
Use the left hole.
AC INPUT terminal block
Terminal block cover
L: Black or brown
N: White or blue
(GND): Green or green and yellow
GND