PFX2512_CE
97
Troubleshooting
App.
The value of voltage / current is not correctly displayed.
Problem
Items to check and corrective actions
Indicates the I/O (Comm) alarm
on the display of the application
softwar.
Indicates when neither the PC nor the PFX2512 are communicated. Please
confirm for the connection status of the system. Moreover, when the power
supply of the PFX2512 is temporarily disconnected, it becomes this alarm. In
case of power failure, please refer to the Help file of the BPChecker3000.
Test Executive
The CC / CV operation has not
been functioned correctly.
Do you connect the other than the battery?
When the connected DUT is other than the battery, there is a possibility that
the operation of the CC / CV control may not be functioned properly.
When at no load (no connection) or the connected battery with the internal
impedance at abnormally high, the CV operation may not be activated
normally.
Do you connect the voltage sensing cable?
Unless otherwise, the operation can not be performed properly without
connecting the voltage sensing cable. Please secure to wire the voltage sensing
cable.
Even though option boards are
installed, the number of
measurement points has not
been expanded.
Are the option boards pushed all the way into the PFX2512?
Have you customized the model ID?
The first time that you use the OP02-PFX after you purchase it, use the
BPChecker3000 to customize the model ID.
See
?
Help
Problem
Items to check and corrective actions
The value has not been
displayed correctly.
Is the item of description for which you wish to display on the front panel been
correctly selected ?
The error of display is large.
Due to the initial temperature drift, the error range may be large immediately
after turning on of the power . Wait until it becomes in stable condition
(approximately 30 minutes), then please confirm it again.
When the dust filter is clogged and the internal temperature rises, the error
range becomes large due to the temperature drift of the measurement circuit
etc.
Clean the dust filter.
When the error becomes large under any condition, it is concerned that the
adjustment may be shifted. Contact Kikusui distributor/agent.
-
The voltage display is unstable.
When the DUT other than battery is connected, or any other device connected
with the DUT, the indication of the voltmeter may not be stable.
-
Is the GND line of the power cable grounded? The indication of the voltmeter
may be fluctuated if it not grounded properly.
The test data varies.
Do you connect the chassis terminal?
It may cause the influence of the noise between the equipments. It is
recommended for the wiring of the chassis terminal in order to prevent the
mis-operation, acquiring the accurate test data, and execution of charge and
discharge test safely.
See
Summary of Contents for PFX2515
Page 10: ...10 PFX2512_CE This page is intentionally blank ...
Page 58: ...58 PFX2512_CE This page is intentionally blank ...
Page 59: ...Specification This chapter contains the PFX2512 specifications and outline drawings ...
Page 86: ...86 PFX2512_CE This page is intentionally blank ...
Page 100: ...100 PFX2512_CE This page is intentionally blank ...
Page 103: ......