Preparation of the DUT (battery) Connection
5
Vo
lt
/
T
h
e
rm
o
m
e
te
r U
n
it
(
O
p
ti
o
n
)
■
Input terminal (Connector MC1.5/10-G : Phoenix Contact)
When the PFX2512 is shipped from the factory, connectors
are attached to the input terminals. If they are damaged or
lost, contact your Kikusui agent or distributor.
Assemble the sensing cable using the wire of size for AWG 24 to 20 (0.20 mm
2
to 0.52 mm
2
).
Temperature measurement (THERMOCOUPLE)
Term No.
Signal
Description
4
Cell T Term4+
Thermocouple connecting terminal 4+
Cell T Term4-
Thermocouple connecting terminal 4-
3
Cell T Term3+
Thermocouple connecting terminal 3+
Cell T Term3-
Thermocouple connecting terminal 3-
2
Cell T Term2+
Thermocouple connecting terminal 2+
Cell T Term2-
Thermocouple connecting terminal 2-
1
Cell T Term1+
Thermocouple connecting terminal 1+
Cell T Term1-
Thermocouple connecting terminal 1-
-
NC
Unused (Reserved)
-
NC
Unused (Reserved)
Voltage measurement (VOLTAGE SENSING)
Term No.
Signal
Description
-
FG
Potential for shielding
4
Cell V Term4+
Cell Voltage sensing connecting terminal 4+
Cell V Term4-
Cell Voltage sensing connecting terminal 4-
3
Cell V Term3+
Cell Voltage sensing connecting terminal 3+
Cell V Term3-
Cell Voltage sensing connecting terminal 3-
2
Cell V Term2+
Cell Voltage sensing connecting terminal 2+
Cell V Term2-
Cell Voltage sensing connecting terminal 2-
1
Cell V Term1+
Cell Voltage sensing connecting terminal 1+
Cell V Term1-
Cell Voltage sensing connecting terminal 1-
-
FG
Potential for shielding
84-61-7910
OP0X
-
PFX
TERM
+
-
+
-
+
-
+
-
1234
THERMO
COUPLE
+
-
+
-
+
-
+
-
1
FG
234
FG
TERM
MAX20V
V
OLT
A
G
E
SENSING
Insert the wire while you press this part down
with a flat-blade screwdriver.
STRIP-GAUGE
7 mm
Insert the wire
with 7 mm of
the covering
removed here.
Yes
No
No
SENSING connector plug
The wire is directly
in contact with the
chassis.
The wire scrap is
in contact with the
chassis.
Summary of Contents for PFX2515
Page 10: ...10 PFX2512_CE This page is intentionally blank ...
Page 58: ...58 PFX2512_CE This page is intentionally blank ...
Page 59: ...Specification This chapter contains the PFX2512 specifications and outline drawings ...
Page 86: ...86 PFX2512_CE This page is intentionally blank ...
Page 100: ...100 PFX2512_CE This page is intentionally blank ...
Page 103: ......