4
Setting up an E5381A Differential Flying Lead Probe Configuration for U4421A Module
36
Keysight U4421A MIPI D-PHY Analyzer and Exerciser Hardware and Probing Guide
Connecting the E5381A Probe to the U4421A Module and DUT
The E5381A probe connects to the DUT via the supplied probing accessories and to the U4421A
module via the U4201A/N2815A cable.
1 Connect the E5381A probe to the U4421A module.
a
Plug the module connector of the U4201A/N2815A cable into the Analysis port on the front
panel of the U4421A module.
b
The module connector of the cable has two thumb screws. Hand-tighten these screws. Do not
over tighten the screws.
c
Connect the other end of the U4201A/N2815A cable to the cable connector component of the
flying lead probe.
2 Connect the E5381A probe to the DUT.
a
Attach the appropriate probe accessories (supplied with the probe) to the DUT. Refer to the
topic
“Connecting the Probe Accessories to the Probe and DUT"
b
Connect the appropriate flying lead cables of the probe to the accessories attached to the DUT
in the previous step. Refer to the topic
“D-PHY Signals to Flying Leads Mapping"
to know which flying lead to connect for probing a particular signal.
Probing Single-ended Signals
To use the E5381A probe for probing single-ended signals
1 Connect the positive side (light grey colored cable) of the flying lead tip to the signal.
2 Connect the negative side (dark grey colored cable) of the flying lead tip to the ground.
Analysis port on U4421A module
NOTE
The U4201A cable has two probe connectors - Odd and Even. If you are using this cable,
then connect the probe to the connector labeled "Odd"on the cable. The probe connector
labeled "Even" on the cable is not used with the U4421A module.
NOTE
You can connect the negative side of the flying lead tip to the ground using the same
connection method that you used for connecting the positive side of the lead to the signal. For
instance, if you used the coaxial tip resistor probe accessory to connect the positive side to the
signal, then you can use the coaxial tip resistor to connect the negative side to the ground.