![Keysight U4421A Hardware And Probing Manual Download Page 33](http://html1.mh-extra.com/html/keysight/u4421a/u4421a_hardware-and-probing-manual_1961143033.webp)
Keysight U4421A MIPI D-PHY Analyzer and Exerciser Hardware and Probing Guide
33
Setting up an E5381A Differential Flying Lead Probe Configuration for U4421A Module
4
You can probe D-Phy Data Lane 0p and Data Lane 0n using two 3-pin connectors laid side-by-side
as follows:
Figure 12
Using two 3-pin headers side-by-side to probe Data Lane 0 p and Data Lane 0 n
There must be 0.1 inch spacing between the 3-pin headers placed side-by-side.
For other D-Phy signals, probe as close as possible to the chip which receives the high-speed data.
The connector dimensions are the same as described for Data Lane 0, but both the positive and
negative sides of the signal will be connected to the same flying lead.