20
U2741A USB Modular 5.5 Digits Digital Multimeter Service Guide
2
Performance Test Calibration
Ohms gain verification test
2- wire ohms configuration:
(
CONFigure:RESistance [<range|MIN|MAX|AUTO>
[,resolution|MIN|MAX|DEF>]])
4- wire ohms configuration:
(
CONFigure:FRESistance [<range|MIN|MAX|AUTO>
[,resolution|MIN|MAX|DEF>]])
1
Select the
Ohms
function.
2
Select each range in the order based on the input
resistance in
Table 2- 5
. Compare the measurement results
with the appropriate test limits shown in the following
table. (Be certain to allow for appropriate source settling.)
[1]
Select NPLC 20 digit resolution.
[2]
The specifications are for 2-wire or 4-wire ohms using the null function in
KMM software. Add 0.2
Ω
error if null function is not used in KMM software.
Q
= Quick performance verification test points
Table 2-5
Ohms gain verification test
Input
Function
[1]
Range
Quick check
Error from nominal
1 year
100
Ω
2-Wire Ohms
and
4-Wire Ohms
100
Ω
±38 m
Ω
[2]
1 k
Ω
1 k
Ω
Q
±350 m
Ω
[2]
10 k
Ω
10 k
Ω
±35
Ω
[2]
100 k
Ω
100 k
Ω
±35
Ω
1 M
Ω
1 M
Ω
±650
Ω
10 M
Ω
10 M
Ω
Q
±25.5 k
Ω
100 M
Ω
100 M
Ω
±2.005 m
Ω
Summary of Contents for U2741A
Page 1: ...Keysight U2741A USB Modular 5 5 Digits Digital Multimeter Service Guide ...
Page 2: ......
Page 3: ......
Page 10: ...VIII U2741A USB Modular 5 5 Digits Digital Multimeter Service Guide ...
Page 14: ...XII U2741A USB Modular 5 5 Digits Digital Multimeter Service Guide Contents ...
Page 15: ...U2741A USB Modular Digital Multimeter Service Guide 1 Specifications Product Specifications 2 ...