Performance Test Calibration
2
U2741A USB Modular 5.5 Digits Digital Multimeter Service Guide
25
Optional AC current performance verification test
These tests are
not
intended to be performed with every
calibration. They are provided as an aid for verifying
additional instrument specifications.
AC current configuration:
(
CONFigure:CURRent:AC [<range|MIN|MAX|AUTO>
[,resolution|MIN|MAX|DEF>]])
1
Select the
AC Current
function.
2
Select each range in the order based on the indicated
input current and frequency. Compare the measurement
results with the appropriate test limits shown in the
Table 2- 10
. (Be certain to allow for appropriate source
settling.)
[1]
Select NPLC 20 digit resolution.
Table 2-10
Optional AC current performance verification test
Current
Input frequency
Function
[1]
Range
Error from nominal 1
year
10 mA
20 Hz
AC Current
10 mA
±0.16 mA
10 mA
45 Hz
10 mA
±0.16 mA
10 mA
1 kHz
10 mA
±0.06 mA
10 mA
10 kHz
10 mA
±0.22 mA
1 A
1 kHz
1 A
±6 mA
100 mA
1 kHz
1 A
±1.5 mA