Performance Test Calibration
2
U2741A USB Modular 5.5 Digits Digital Multimeter Service Guide
17
[1]
Select NPLC 20 digit resolution.
[2]
The specifications are for 2-wire or 4-wire ohms using the null function in KMM
software. Add 0.2
Ω
error if null function is not used in KMM software.
Q
= Quick performance verification test points
Short
DC Volts
100 mV
±8 µV
Short
1 V
Q
±50 µV
Short
10 V
±0.5 mV
Short
100 V
±5 mV
Short
300 V
±15 mV
Short
2-Wire Ohms
and
4-Wire Ohms
100
Ω
±8 m
Ω
[2]
Short
1 k
Ω
±50 m
Ω
[2]
Short
10 k
Ω
Q
±500 m
Ω
[2]
Short
100 k
Ω
±5
Ω
Short
1 M
Ω
±50
Ω
Short
10 M
Ω
±500
Ω
Short
100 M
Ω
±5 k
Ω
Table 2-2
Zero offset verification test
Input
Function
[1]
Range
Quick check
Error from nominal
1 year