Troubleshooting
1
Keysight RP7900 Series Advanced Service Guide
35
– For the control board, measure the V
mon
, V
loc
, and I
mon
signals that goes
into the board and verify that the values are accurate. Send the following
SCPI commands via the remote interface:
DIAG:MEAS:VOLT:LOC?
– measures the local voltage from the aux ADC
DIAG:MEAS:AUX:CURR?
- measures the current from the aux ADC
DIAG:MEAS:AUX:POWER?
- measures the power from the aux ADC
DIAG:MEAS:MAIN:POW:LOC?
- measures the local power from the main ADC
8
Refer to the troubleshooting flow chart in
for further details.
SCPI Command
Expected return value
[a]
[a] Instrument setting - Voltage: 20 V, Current: 1 A
Expected return value
[b]
[b] Instrument setting - Voltage: 500 V, Current: 8 A
DIAG:MEAS:VOLT:LOC?
21.45 V
499.72 V
DIAG:MEAS:AUX:CURR?
1.00 A
8.007 A
DIAG:MEAS:AUX:POWER?
20.11 W
4003.14 W
DIAG:MEAS:MAIN:POW:L
OC?
0.0806 W
16.013 W