Troubleshooting
28
Keysight RP7900 Series Advanced Service Guide
110
Wrong V+I
The instrument was unable to set the correct voltage or current value.
Refer to the
114
Wrong status
An incorrect status function has been reported.
116
Locked out by internal switch setting
This function has been locked out by an internal switch.
Refer to the
Operating and
Service Guide
117
Calibration error
A calibration error has occurred. Do not save calibration constants. Try
re-calibrating the unit.
Refer to the Calibration
procedure in the
Operating and
Service Guide
200
Hard ware error channel <1>
A hardware error has occurred on the output.
Refer to the
201
Invalid configuration
An invalid parallel or SDS configuration is not allowed.
Check hardware model number
and firmware revision
202
Self-test fail
A self test failure has occurred.
Refer to the
203
Compatibility function not implemented
The requested compatibility function is not available.
Check firmware revision
204
NVRAM checksum error
A checksum error has occurred in the instrument’s nonvolatile random access
memory.
Check the P600 and constellation
flash memory or EEPROM
205
NVRAM full
The nonvolatile random access memory of the instrument is full.
206
File not found
The internal calibration file or the internal channel attribute file was not found in
NVRAM.
Check the DPM
207
Cal file version error
The calibration file was written or read using old firmware. The firmware must be
updated.
Update the firmware revision
208
Running backup firmware
The instrument is presently running the backup (previous) version of the firmware.
Check the P600 firmware
programming
Error
code Description
Potential remedy