Disassembly
2
Keysight RP7900 Series Advanced Service Guide
63
NOTE
– The FFC cable is secured to the constellation and P600 assembly connector
via glue at both sides. If you need to replace the constellation and P600
assembly, carefully remove the glue first before you remove the FFC cable.
– When reconnecting the FFC cable to the constellation and P600 assembly
reapply the glue (on both sides of the cable) using a glue gun to secure the
FFC cable to the constellation and P600 assembly. Ensure that the FFC cable
is inserted properly before applying the glue.