6-2
Installation and Service Guide N5251-90001
Performance Tests and Checks
System Preparation and Analyzer Warm Up
PNA Series Microwave Network Analyzer System
N5251A
System Preparation and Analyzer Warm Up
To achieve the maximum system stability, allow the analyzer to warm up for
at least 90 minutes.
Chapter 3 , “System Installation”
system. After you press the Preset button, the PNA should display a frequency range of 10 MHz
to 110 GHz.
Torquing Connections
All connections made during the System Check and System Performance Verification
procedures should be carefully torqued using the tools provided in the 85059A kit.
Long Term Storage of Test Results
It is recommended that you store results from the System Check and the System Performance
Verification procedures for future reference. Prior results can be useful when evaluating
changes in system performance. After completing a successful test process that accurately
represents the system
performance, store the test result files in the directory
D:\sysver
results
.
It may be necessary to create this directory if it does not already exist. Create a new
subdirectory when a new set of test results is stored. Use the current date as the name for the
subdirectory. Example:
D:\sysver results\201
6
April 15\.
Protect Against Electrostatic Discharge (ESD)
This is important. If not properly protected against, electrostatic discharge can seriously
damage your analyzer, resulting in costly repair.
To reduce the chance of electrostatic discharge, follow all of the
“Electrostatic Discharge Protection” on
, for all of the procedures in this chapter.
Summary of Contents for N5227A
Page 4: ...iii Installation and Service Guide N5251 90001...
Page 7: ...Installation and Service Guide N5251 90001 1 1 1 Safety and Regulatory Information...
Page 15: ...Installation and Service Guide N5251 90001 2 1 2 System Description...
Page 25: ...Installation and Service Guide N5251 90001 3 1 3 System Installation...
Page 41: ...Installation and Service Guide N5251 90001 4 1 4 System Specifications...
Page 49: ...Installation and Service Guide N5251 90001 5 1 5 Maintenance and Support...
Page 53: ...Installation and Service Guide N5251 90001 6 1 6 Performance Tests and Checks...
Page 69: ...Installation and Service Guide N5251 90001 7 1 7 Replaceable Parts...