Installation and Service Guide N5251-90001
6-9
PNA Series Microwave Network Analyzer System
N5251A
Performance Tests and Checks
System Performance Verification
Figure 6-4
System Verification Configure Tab
9. Select the Run tab to return to the previous window.
The remaining instructions assume Port 1 and Port 2 are being
verified. Modify the instructions appropriately for Port 3 and Port 4.
10. Click Run.
11. Enter the serial number for the 85059A kit. The other numbers are optional.
12. Click Continue.
13. As prompted, install the 1.0 mm female to female cable (8.8 cm, Keysight part number
11500-60001) onto the right test head (Port 2). See
. This cable is
part of the 85059A 1.0 mm Calibration and Verification kit and is considered to be a test
port cable.
Summary of Contents for N5227A
Page 4: ...iii Installation and Service Guide N5251 90001...
Page 7: ...Installation and Service Guide N5251 90001 1 1 1 Safety and Regulatory Information...
Page 15: ...Installation and Service Guide N5251 90001 2 1 2 System Description...
Page 25: ...Installation and Service Guide N5251 90001 3 1 3 System Installation...
Page 41: ...Installation and Service Guide N5251 90001 4 1 4 System Specifications...
Page 49: ...Installation and Service Guide N5251 90001 5 1 5 Maintenance and Support...
Page 53: ...Installation and Service Guide N5251 90001 6 1 6 Performance Tests and Checks...
Page 69: ...Installation and Service Guide N5251 90001 7 1 7 Replaceable Parts...